Suppr超能文献

计算机X线摄影系统中图像质量与曝光指数及处理协议的相关性

Correlation of image quality with exposure index and processing protocol in a computed radiography system.

作者信息

Tsalafoutas I A, Blastaris G A, Moutsatsos A S, Chios P S, Efstathopoulos E P

机构信息

Medical Physics Department, Agios Savvas Hospital, 171 Alexandras Avenue, 115 22 Athens, Greece.

出版信息

Radiat Prot Dosimetry. 2008;130(2):162-71. doi: 10.1093/rpd/ncm493. Epub 2008 Feb 2.

Abstract

The correlation of image quality with the exposure index (EI) and the processing protocol was investigated in a Kodak computed radiography (CR) system using clinical radiographs and a water phantom containing an aluminium and a copper step-wedge. The phantom was exposed to different dose levels and the acquired images were processed using four clinical protocols. The quality of these images was evaluated in terms of image brightness, contrast and noise. In clinical radiographs, there was no straightforward correlation of image quality with EI. In phantom images, higher EI values improved contrast and reduced noise but after a point this improvement does not justify the implied increase in patient dose. Image brightness, contrast and noise were also strongly dependent on the processing protocol. To obtain the images of satisfactory quality with the Kodak CR system, a dose slightly higher than those used in 400 relative speed screen-film systems and a processing protocol designated for the specific radiographic examination are required.

摘要

在柯达计算机X线摄影(CR)系统中,使用临床X线片以及含有铝和铜阶梯楔形块的水模体,研究了图像质量与曝光指数(EI)和处理协议之间的相关性。该模体接受不同剂量水平的照射,采集的图像使用四种临床协议进行处理。这些图像的质量通过图像亮度、对比度和噪声进行评估。在临床X线片中,图像质量与EI之间没有直接的相关性。在模体图像中,较高的EI值可改善对比度并降低噪声,但超过某一点后,这种改善并不能证明患者剂量的隐含增加是合理的。图像亮度、对比度和噪声也强烈依赖于处理协议。为了用柯达CR系统获得质量令人满意的图像,需要比400相对速度增感屏-胶片系统所使用的剂量略高的剂量以及针对特定X线检查指定的处理协议。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验