Jeevanantham Vinodh, Levine Ethan, Budzikowski Adam S, Shah Abrar H, Daubert James P
Unity Hospital, Rochester, New York, USA.
Pacing Clin Electrophysiol. 2008 Mar;31(3):375-7. doi: 10.1111/j.1540-8159.2008.01000.x.
We report a rare case of inadvertent reversal of the defibrillator shock coils during generator change which then resulted in both intermittent, recurrent "noise" on both the rate sensing and the shock electrogram leads and multiple inappropriate shocks.