Chinushi Masaomi, Hosaka Yukio, Ikarashi Noboru, Iijima Kenichi, Furushima Hiroshi, Aizawa Yoshifusa
School of Health Science, Niigata University School of Medicine, 2-746 Asahimachi, Chuou-ku, Niigata 951-8518, Japan.
Europace. 2008 Jun;10(6):738-40. doi: 10.1093/europace/eun095. Epub 2008 Apr 4.
A 62-year-old man was afflicted with implantable cardioverter defibrillator (ICD) shocks during sinus rhythm. Stored ICD data revealed that sensing of noise due to fracture of the ventricular lead triggered the delivery of shocks. Since the lead fracture developed suddenly, it is suggested that close, early attention should be paid to the potential of such events during follow-up of ICD leads.
一名62岁男性在窦性心律时遭受植入式心律转复除颤器(ICD)电击。ICD存储数据显示,心室导线断裂产生的噪音感知触发了电击的发放。由于导线断裂突然发生,提示在ICD导线随访期间应密切、早期关注此类事件的可能性。