Suppr超能文献

视网膜厚度分析仪与GDx VCC扫描激光偏振仪、Stratus OCT光学相干断层扫描仪及海德堡视网膜断层扫描仪II共焦扫描激光检眼镜之间的关系。

Relationship between the retinal thickness analyzer and the GDx VCC scanning laser polarimeter, Stratus OCT optical coherence tomograph, and Heidelberg retina tomograph II confocal scanning laser ophthalmoscopy.

作者信息

Ma Kyoung Tak, Lee Sang Hyup, Hong Samin, Park Kyoung Soo, Kim Chan Yun, Seong Gong Je, Hong Young Jae

机构信息

Siloam Eye Hospital, Seoul, Korea.

出版信息

Korean J Ophthalmol. 2008 Mar;22(1):10-7. doi: 10.3341/kjo.2008.22.1.10.

Abstract

PURPOSE

To assess the relationship between the retinal thickness analyzer (RTA) parameters, and those of the GDx VCC scanning laser polarimeter (GDx VCC), Stratus OCT optical coherence tomography (Stratus OCT), and Heidelberg retinal tomograph II confocal scanning laser ophthalmoscopy (HRT II).

METHODS

Twenty-nine primary open-angle glaucoma patients were retrospectively included in this study. Measurements were obtained using the RTA, GDx VCC, Stratus OCT, and HRT II. We calculated the correlation coefficients between the parameters of RTA and those of the other studies.

RESULTS

Among the optic disc parameters of RTA, the cup volume was best correlated with Stratus OCT (R=0.780, p<0.001) and HRT II (R=0.896, p<0.001). Among the posterior pole retinal thickness parameters, the posterior pole abnormally thin area (PPAT) of the RTA and the inferior average of the GDx VCC were best correlated (R=-0.596, p=0.001). The PPAT of the RTA and the inferior maximum of the Stratus OCT were best correlated (R=-0.489, p=0.006). The perifoveal minimum thickness (PFMT) of the RTA and the cup shape measurement of the HRT II were best correlated (R=-0.565, p=0.004).

CONCLUSIONS

Many RTA optic disc parameters were significantly correlated with those of the Stratus OCT and HRT II. The RTA posterior pole retinal thickness parameters were significantly correlated with those of the GDx VCC, Stratus OCT and HRT II. The RTA optic disc and posterior pole retinal thickness parameters may be valuable in the diagnosis of glaucoma.

摘要

目的

评估视网膜厚度分析仪(RTA)参数与GDx VCC扫描激光偏振仪(GDx VCC)、Stratus OCT光学相干断层扫描(Stratus OCT)以及海德堡视网膜断层扫描仪II共焦扫描激光检眼镜(HRT II)参数之间的关系。

方法

本研究回顾性纳入了29例原发性开角型青光眼患者。使用RTA、GDx VCC、Stratus OCT和HRT II进行测量。我们计算了RTA参数与其他研究参数之间的相关系数。

结果

在RTA的视盘参数中,杯状容积与Stratus OCT(R = 0.780,p < 0.001)和HRT II(R = 0.896,p < 0.001)的相关性最佳。在黄斑后极视网膜厚度参数中,RTA的黄斑后极异常变薄区域(PPAT)与GDx VCC的下方平均值相关性最佳(R = -0.596,p = 0.001)。RTA的PPAT与Stratus OCT的下方最大值相关性最佳(R = -0.489,p = 0.006)。RTA的中心凹周围最小厚度(PFMT)与HRT II的杯状形态测量相关性最佳(R = -0.565,p = 0.004)。

结论

许多RTA视盘参数与Stratus OCT和HRT II的参数显著相关。RTA黄斑后极视网膜厚度参数与GDx VCC、Stratus OCT和HRT II的参数显著相关。RTA视盘和黄斑后极视网膜厚度参数在青光眼诊断中可能具有重要价值。