Suzuki Shigeru, Furui Shigeru, Isshiki Takaaki, Kozuma Ken, Koyama Yutaka, Ochiai Masahiko, Asakura Yasushi
Department of Radiology, Teikyo University School of Medicine, Kaga, Itabashi-ku, Tokyo, Japan.
Catheter Cardiovasc Interv. 2008 May 1;71(6):792-8. doi: 10.1002/ccd.21481.
The purpose of this research is to assess the patient's entrance skin dose (ESD) during percutaneous coronary intervention (PCI) for chronic total occlusion (CTO), and discuss methods to reduce the maximum ESDs.
Only a few reports are available on the methods to reduce patients' maximum ESD during the procedures.
This study included consecutive 30 patients who underwent PCI procedures for CTO in the three institutions. Pearson correlation test was employed to determine the relationship between total fluoroscopic time (TFT) and the maximum ESD, dose area product (DAP) value, and the maximum ESD in each institution.
There were significant correlations between the TFT and maximum ESD (Institution 1: P = 0.000410, Institution 2: P = 0.000525), and between the DAP and the maximum ESD (Institution 2: P < 0.0001). In Institution 1, TFT of 60 min was set as the upper limit, and the maximum ESDs were controlled within 7 Gy. In Institution 2, the angiographic unit was a biplane system, and two skin sites were exposed, corresponding to the angulation of each X-ray tube. In Institution 3, the interventionalist changed the beam angulations frequently by several degrees during the procedures, and the maximum ESD was controlled within 3 Gy even during procedures with a TFT of more than 1 hr.
The TFT and DAP, the latter of which is preferable, are useful to estimate the maximum ESD. Limiting the TFT or DAP, or changing the beam angulations is important to control ESD during prolonged procedures.
本研究旨在评估慢性完全闭塞病变(CTO)经皮冠状动脉介入治疗(PCI)期间患者的入射皮肤剂量(ESD),并探讨降低最大ESD的方法。
关于在手术过程中降低患者最大ESD的方法,仅有少数报告。
本研究纳入了在三家机构接受CTO-PCI手术的连续30例患者。采用Pearson相关检验确定总透视时间(TFT)与最大ESD、剂量面积乘积(DAP)值以及各机构最大ESD之间的关系。
TFT与最大ESD之间存在显著相关性(机构1:P = 0.000410,机构2:P = 0.000525),DAP与最大ESD之间也存在显著相关性(机构2:P < 0.0001)。在机构1中,将60分钟的TFT设定为上限,最大ESD控制在7 Gy以内。在机构2中,血管造影设备为双平面系统,两个皮肤部位暴露,对应于每个X射线管的角度。在机构3中,介入医生在手术过程中频繁将束角改变几度,即使在TFT超过1小时的手术中,最大ESD也控制在3 Gy以内。
TFT和DAP(后者更可取)有助于估计最大ESD。在长时间手术中,限制TFT或DAP,或改变束角对于控制ESD很重要。