Tauhata Luiz, Iwahara Akira, de Oliveira Antonio E, Rezende Eduarda Alexandre, Delgado José Ubiratan, da Silva Carlos José, Dos Santos Joyra A, Nícoli Ieda G, Alabarse Frederico G, Xavier Ana Maria
Laboratório Nacional de Metrologia das Radiações Ionizantes, Instituto de Radioproteção e Dosimetria, Comissão Nacional de Energia Nuclear, Rio de Janeiro, Brazil.
Appl Radiat Isot. 2008 Jun-Jul;66(6-7):981-7. doi: 10.1016/j.apradiso.2008.02.045. Epub 2008 Feb 13.
Proficiency tests were applied to assess the performance of 74 nuclear medicine services in activity measurements of (131)I, (123)I, (99)Tc(m), (67)Ga and (201)Tl. These tests produced 913 data sets from comparison programmes promoted by the National Laboratory for Ionizing Radiation Metrology (LNMRI) from 1999 to 2006. The data were evaluated against acceptance criteria for accuracy and precision and assigned as Acceptable or Not acceptable accordingly. In addition, three other statistical parameters were used as complementary information for performance evaluation which related to normative requirements and to radionuclide calibrators. The results have shown a necessity to improve quality control procedures and unsatisfactory performances of radionuclide calibrators, which incorporate Geiger-Müller detectors.