Comparison of the activity measurements in nuclear medicine services in the Brazilian northeast region.

作者信息

de Farias Fragoso Maria da Conceição, de Albuquerque Antônio Morais, de Oliveira Mércia L, de Lima Fabiana Farias, Barreto Flávio Chiappetta Paes, de Andrade Lima Ricardo

机构信息

Centro Regional de Ciências Nucleares do Nordeste (CRCN-NE), Comissão Nacional de Energia Nuclear (CNEN), Av. Prof. Luiz Freire, 200, CEP 50740-540, Recife/PE, Brazil.

出版信息

Appl Radiat Isot. 2013 Dec;82:36-44. doi: 10.1016/j.apradiso.2013.07.014. Epub 2013 Jul 24.

Abstract

The Northeastern Regional Centre for Nuclear Sciences (CRCN-NE), National Nuclear Energy Commission, has organized for the first time in nuclear medicine services (NMSs) in the Brazilian northeast region a comparison of activity measurements for (99m)Tc, (131)I, (67)Ga, (201)Tl and (57)Co. This tool is widely utilized to evaluate not only the accuracy of radionuclide calibrators, but also the competence of NMSs to measure the activity of the radiopharmaceuticals and the performance of the personnel involved in these measurements. The comparison results showed that 90% of the results received from participants are within the ±10% limit established by the Brazilian Norm.