Abstract

StressEraser is a commercially marketed biofeedback device designed to enhance heart rate variability. StressEraser makes its internal calculations on beat-to-beat measures of finger pulse intervals. However, the accuracy and precision of StressEraser in quantifying interbeat intervals using finger pulse intervals has not been evaluated against standard laboratory equipment using R-R intervals. Accuracy was assessed by simultaneously recording interbeat intervals using StressEraser and a standard laboratory ECG system. The interbeat intervals were highly correlated between the systems. The average deviation in interbeat interval recordings between the systems was approximately 6 ms. Moreover, correlations approached unity between the systems on estimates of heart period, heart rate, and heart rate variability. Feedback from StressEraser is based on an interbeat time series that provides sufficient information to provide an excellent estimate of the dynamic changes in heart rate and heart rate variability. The slight variations between StressEraser and the laboratory equipment in quantifying heart rate and heart rate variability are due to features related to monitoring heart rate with finger pulse: (1) a lack in precision in the peak of the finger pulse relative to the clearly defined inflection point in the R-wave, and (2) contribution of variations in pulse transit time.

摘要

应激消除器是一种商业销售的生物反馈设备,旨在提高心率变异性。应激消除器根据逐搏测量的手指脉搏间期进行内部计算。然而,应激消除器使用手指脉搏间期量化心搏间期的准确性和精密度尚未与使用R-R间期的标准实验室设备进行对比评估。通过使用应激消除器和标准实验室心电图系统同时记录心搏间期来评估准确性。两个系统的心搏间期高度相关。两个系统在心搏间期记录中的平均偏差约为6毫秒。此外,在心动周期、心率和心率变异性估计方面,两个系统之间的相关性接近1。应激消除器的反馈基于一个心搏时间序列,该序列提供了足够的信息来很好地估计心率和心率变异性的动态变化。应激消除器与实验室设备在量化心率和心率变异性方面的细微差异是由于与用手指脉搏监测心率相关的特征所致:(1)手指脉搏峰值相对于R波中明确定义的拐点缺乏精度,以及(2)脉搏传输时间变化的影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验