Xiang Ping, Shen Min, Zhuo Xian-Yi
[Shanghai Key Laboratory of Forensic Medicine, Institute of Forensic Sciences, Ministry of Justice, PR China, Shanghai 200063, China]
Fa Yi Xue Za Zhi. 2008 Feb;24(1):60-5.
It is imperative that any newly established bio-analytical method is validated thoroughly, using standardised parameters. The purpose of this article is to provide recommendations on how to validate a new bio-analytical method. Based on author's personal experience and some interesting discussion points from the conference of "International Association of Forensic Toxicologists" in 2007, the authors propose these essential requirements for validating a new analytical method. The key parameters of method validation include selectivity, linearity, accuracy, precision, LOD (limit of detection), LLOQ (the lower limit of quantitation), stability and the extraction recovery. For any bio-analytical method using LC-MSn (Liquid chromatography-mass spectrometry), studies of matrix effect should also be included in addition of the above parameters.
任何新建立的生物分析方法都必须使用标准化参数进行全面验证。本文旨在就如何验证一种新的生物分析方法提供建议。基于作者的个人经验以及2007年“国际法医毒理学家协会”会议上的一些有趣讨论点,作者提出了验证新分析方法的这些基本要求。方法验证的关键参数包括选择性、线性、准确度、精密度、检测限(LOD)、定量下限(LLOQ)、稳定性和提取回收率。对于任何使用液相色谱-质谱联用(LC-MSn)的生物分析方法,除上述参数外,还应包括基质效应研究。