Campbell Charles E
Advanced Medical Optics, Santa Ana, Calif, USA.
J Refract Surg. 2008 Mar;24(3):308-11. doi: 10.3928/1081597X-20080301-15.
To investigate whether wavefront measurements taken in eyes that have implanted multifocal intraocular lenses (IOLs) are reliable when creating laser refractive surgical treatments to remove residual refractive error from these eyes.
A specially designed fluid-filled model eye holding a multifocal IOL was refracted using a commercial wavefront eye refractor. The wavefront findings were then compared to the expected refractive error. The fluid-filled model was designed to match the optical characteristics of a human eye and to hold the IOL at the same distance from the cornea as in a human eye. Lenses tested were AMO Tecnis ZM900 series (diffractive design, +25.00 D, +4.00 D add), AMO ReZoom NXG1 series (refractive design, 25.00 D, +4.00 D add), and AMO 811E (diffractive with aspheric design 25.00 D, +4.00 add). The visible pupil diameter was 5.75 mm.
Diffractive bifocal IOLs were found to create spot doubling in the Shack-Hartmann sensor image as expected from theoretical considerations. However, the residual sphere power and cylinder power reported were those expected. In addition, the higher order aberrations reported were reasonable in light of expected results. Refractive multifocal IOLs were found to create distorted Shack-Hartmann images as expected. The residual sphere power and cylinder power reported were not those expected nor were the higher order aberrations. In addition, the higher order values were sensitive to the position of the eye with respect to the refractor.
Wavefront measurements from eyes with diffractive IOLs can be used, with caution, to plan laser refractive treatments. Wavefront measurements from eyes with refractive multifocal IOLs should not be used to plan post-implant laser refractive correction.
探讨在为植入多焦点人工晶状体(IOL)的眼睛进行激光屈光手术以消除其残余屈光不正时,所进行的波前测量是否可靠。
使用商用波前眼屈光仪对一个装有多焦点IOL的特殊设计的充液模型眼进行屈光测量。然后将波前测量结果与预期的屈光不正进行比较。该充液模型旨在匹配人眼的光学特性,并将IOL置于与人眼中相同的距角膜距离处。测试的晶状体包括AMO Tecnis ZM900系列(衍射设计,+25.00 D,+4.00 D附加光度)、AMO ReZoom NXG1系列(折射设计,25.00 D,+4.00 D附加光度)以及AMO 811E(非球面衍射设计,25.00 D,+4.00附加光度)。可见瞳孔直径为5.75毫米。
如理论考虑所预期的那样,发现衍射双焦点IOL在夏克-哈特曼传感器图像中产生光斑加倍现象。然而,所报告的残余球镜度和柱镜度为预期值。此外,根据预期结果,所报告的高阶像差是合理的。发现折射多焦点IOL如预期那样产生扭曲的夏克-哈特曼图像。所报告的残余球镜度和柱镜度并非预期值,高阶像差也不是。此外,高阶值对眼睛相对于屈光仪的位置敏感。
对于植入衍射IOL的眼睛,波前测量结果可谨慎用于规划激光屈光治疗。对于植入折射多焦点IOL的眼睛,波前测量结果不应被用于规划植入后的激光屈光矫正。