Effect of soft contact lenses on optical measurements of axial length and keratometry for biometry in eyes with corneal irregularities.

作者信息

Lewis Jennifer R, Knellinger Andrea E, Mahmoud Ashraf M, Mauger Thomas F

机构信息

Department of Ophthalmology, Havener Eye Institute, The Ohio State University, Columbus, Ohio 43210, USA. mailto:

出版信息

Invest Ophthalmol Vis Sci. 2008 Aug;49(8):3371-8. doi: 10.1167/iovs.07-1247. Epub 2008 Apr 25.

Abstract

PURPOSE

To assess the repeatability and reliability of IOLMaster (Carl Zeiss Meditec, Inc., Dublin, CA) axial length and keratometry measurements (K readings) with a soft contact lens on normal eyes. The method is designed for eyes with corneal irregularities or after endothelial keratoplasty.

METHODS

Biometry was performed on 20 healthy right eyes of volunteer subjects with mean age, 27.3 +/- 4.9 years; axial length, 24.77 +/- 1.04 mm; and K reading, 43.48 +/- 1.69 D. Axial length and keratometry were measured and repeated with -0.5 D SofLens38 (Bausch & Lomb, Rochester, NY) and Acuvue2 (Johnson & Johnson, New Brunswick, NJ) soft contact lenses. Repeatability and reliability were evaluated. Contact lens thickness was measured directly by corneal optical coherence tomography (OCT).

RESULTS

Axial lengths increased 59 +/- 10 microm with SofLens38 and 134 +/- 13 microm with Acuvue2, and these changes correlated with the OCT contact lens thicknesses (P = 0.995). The axial length variability remained constant (P = 0.18), measuring 24 +/- 10 microm for SofLens38 and 23 +/- 8 microm for Acuvue2 compared with 20 +/- 7 microm with no lens. K readings of 43.08 +/- 1.66 D with SofLens38, 42.79 +/- 1.57 D with Acuvue2, and 43.48 +/- 1.69 D with no lens corresponded to differences of -0.40 +/- 0.12 D with SofLens38 and -0.69 +/- 0.19 D with Acuvue2. The K-reading variability increased slightly from 0.04 to 0.09 D with either lens.

CONCLUSIONS

Low-power soft contact lenses enable reliable and repeatable IOLMaster axial length and K-reading measurements. Correcting for the measurable lens thickness and lens effects, a <0.5-D error in the Sanders-Retzlaff-Kraff (SRK) II power formula is predicted.

摘要

目的

评估IOLMaster(卡尔蔡司医疗技术公司,加利福尼亚州都柏林)在正常眼睛上使用软性接触镜时测量眼轴长度和角膜曲率(K值读数)的可重复性和可靠性。该方法适用于角膜不规则或接受内皮角膜移植术后的眼睛。

方法

对20名志愿者的健康右眼进行生物测量,志愿者平均年龄为27.3±4.9岁;眼轴长度为24.77±1.04mm;K值读数为43.48±1.69D。分别使用-0.5D的SofLens38(博士伦公司,纽约州罗切斯特)和Acuvue2(强生公司,新泽西州新不伦瑞克)软性接触镜测量并重复测量眼轴长度和角膜曲率。评估其可重复性和可靠性。通过角膜光学相干断层扫描(OCT)直接测量接触镜厚度。

结果

使用SofLens38时眼轴长度增加59±10μm,使用Acuvue2时增加134±13μm,这些变化与OCT测量的接触镜厚度相关(P = 0.995)。眼轴长度的变异性保持不变(P = 0.18),使用SofLens38时为24±10μm,使用Acuvue2时为23±8μm,而不使用镜片时为20±7μm。使用SofLens38时K值读数为43.08±1.66D,使用Acuvue2时为42.79±1.57D,不使用镜片时为43.48±1.69D,对应于使用SofLens38时差值为-0.40±0.12D,使用Acuvue2时差值为-0.69±0.19D。使用任何一种镜片时,K值读数的变异性从0.04D略有增加至0.09D。

结论

低度数软性接触镜可实现可靠且可重复的IOLMaster眼轴长度和K值读数测量。校正可测量的镜片厚度和镜片效应后,预测桑德斯-雷茨拉夫-克拉夫(SRK)II屈光力公式中的误差<0.5D。