To compare the repeatability and agreement in biometric measurements using Spectral Domain Anterior Segment OCT (AS-OCT, REVO-NX, Optopol) and Scheimpflug tomography (Pentacam-AXL, Oculus) in keratoconus.
Prospective case series at a university hospital tertiary center. Axial length (AL), anterior chamber depth (ACD), central corneal thickness (CCT), and thinnest corneal thickness (TCT) were measured using both devices in patients with keratoconus. Three groups were analyzed: eyes with no prior crosslinking or contact lens wear (Group A), eyes with prior crosslinking (Group B), and eyes with prior contact lens wear (Group C). Repeatability and agreement of measurements were analyzed.
The study comprised of 214 eyes of 157 subjects. In Group A (n = 95 eyes), Group B (n = 86 eyes), and Group C (n = 33 eyes), intraclass correlation coefficient (ICC) was higher than 0.90 for all examined parameters, except for ACD readings in Group A with the REVO-NX (ICC = 0.83). Differences in ACD, TCT, and CCT were significantly different between the two devices for Groups A, B and C (p<0.05). AL measurements differed significantly in Groups A and B (p<0.05) but not in Group C (p = 0.18). Repeatability did not vary significantly between Groups A, B, or C in any parameter with both devices (p>0.05). There was poor agreement between the two devices across all parameters (p<0.05).
Both devices demonstrated good repeatability but poor agreement across AL, ACD, CCT and TCT measurements. There was no significant difference in repeatability in virgin eyes compared to eyes with prior crosslinking or contact lens wear, however, the interchangeable use of the two devices is not recommended.
比较眼前节相干光断层扫描(AS-OCT,REVO-NX,Optopol)和Scheimpflug 断层扫描(Pentacam-AXL,Oculus)在圆锥角膜中的生物测量重复性和一致性。
这是一家大学医院的三级中心的前瞻性病例系列研究。使用这两种设备在圆锥角膜患者中测量眼轴(AL)、前房深度(ACD)、中央角膜厚度(CCT)和最薄角膜厚度(TCT)。分析了三组:未进行交联或接触镜佩戴的眼(A 组)、已进行交联的眼(B 组)和已进行接触镜佩戴的眼(C 组)。分析了测量的重复性和一致性。
该研究共纳入 157 名患者的 214 只眼。在 A 组(n = 95 只眼)、B 组(n = 86 只眼)和 C 组(n = 33 只眼)中,除了使用 REVO-NX 时 A 组的 ACD 读数(ICC = 0.83)外,所有检查参数的组内相关系数(ICC)均高于 0.90。对于 A、B 和 C 组,两种设备测量的 ACD、TCT 和 CCT 差异均有统计学意义(p<0.05)。A 组和 B 组的 AL 测量值差异有统计学意义(p<0.05),但 C 组无差异(p = 0.18)。对于任何参数,两种设备在 A、B 或 C 组中的重复性均无显著差异(p>0.05)。两种设备在所有参数上的一致性均较差(p<0.05)。
两种设备在 AL、ACD、CCT 和 TCT 测量方面均表现出良好的重复性和较差的一致性。与已进行交联或接触镜佩戴的眼相比,未经交联或接触镜佩戴的眼的重复性无显著差异,但不建议两种设备交替使用。