Rocha-Cabral Renata Maciel, Mendes Fausto Medeiros, Miura Fabiana, Ribeiro Adriana Da Costa, Braga Mariana Minatel, Zezell Denise Maria
Centro de Lasers e Aplicações, Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, São Paulo, Brazil.
Acta Odontol Scand. 2008 Apr;66(2):122-7. doi: 10.1080/00016350802017484.
To evaluate the influence of probe tip autoclaving and depleted alkaline batteries on laser fluorescence (LF-DIAGNOdent) device performance.
One-hundred-and-twenty occlusal sites were analyzed with an LF device in nine conditions: No autoclaved probe and probe after 5, 10, 15, 20, 25, 30, 40, and 50 autoclaving cycles. Subsequently, the performance of the device was analyzed with: New batteries, batteries with 1.49/1.39 V, 1.38/1.37 V, 1.36/1.34 V, 1.33/1.32 V, and lower than 1.32 V. LF values, sensitivity, specificity, accuracy, and area under the ROC curve were compared.
In the probe study, sensitivity was lower after 50 sterilizing cycles, though specificity was higher than the assessment performed using a new tip. In the batteries study, specificity was higher for depleted batteries, but LF performance did not differ significantly among the groups.
Batteries do not significantly influence device performance, but consecutive sterilization of probes in autoclave alters readings, downgrading its performance.
评估探针尖端高压灭菌和碱性电池电量耗尽对激光荧光(LF-DIAGNOdent)设备性能的影响。
使用LF设备在九种条件下对120个咬合面进行分析:未高压灭菌的探针以及经过5、10、15、20、25、30、40和50次高压灭菌循环后的探针。随后,使用以下电池分析设备性能:新电池、电压为1.49/1.39 V、1.38/1.37 V、1.36/1.34 V、1.33/1.32 V以及低于1.32 V的电池。比较LF值、灵敏度、特异性、准确性和ROC曲线下面积。
在探针研究中,经过50次灭菌循环后灵敏度较低,不过特异性高于使用新尖端进行的评估。在电池研究中,电量耗尽的电池特异性较高,但各组之间的LF性能无显著差异。
电池对设备性能影响不显著,但在高压灭菌器中对探针进行连续灭菌会改变读数,降低其性能。