迈向通用的二级质谱图库——跨越十一种不同质谱仪的二级质谱图重现性

Towards a universal product ion mass spectral library - reproducibility of product ion spectra across eleven different mass spectrometers.

作者信息

Hopley Chris, Bristow Tony, Lubben Anneke, Simpson Alec, Bull Elaine, Klagkou Katerina, Herniman Julie, Langley John

机构信息

LGC, Queens Road, Teddington TW11 0LY, UK.

出版信息

Rapid Commun Mass Spectrom. 2008 Jun;22(12):1779-86. doi: 10.1002/rcm.3545.

Abstract

Product ion spectra produced by collision-induced dissociation (CID) in tandem mass spectrometry experiments can differ markedly between instruments. There have been a number of attempts to standardise the production of product ion spectra; however, a consensus on the most appropriate approach to the reproducible production of spectra has yet to be reached. We have previously reported the comparison of product ion spectra on a number of different types of instruments - a triple quadrupole, two ion traps and a Fourier transform ion cyclotron resonance mass spectrometer (Bristow AWT, Webb KS, Lubben AT, Halket JM. Rapid Commun. Mass Spectrom. 2004; 18: 1). The study showed that a high degree of reproducibility was achievable. The goal of this study was to improve the comparability and reproducibility of CID product ion mass spectra produced in different laboratories and using different instruments. This was carried out experimentally by defining a spectral calibration point on each mass spectrometer for product ion formation. The long-term goal is the development of a universal (instrument independent) product ion mass spectral library for the identification of unknowns. The spectra of 48 compounds have been recorded on eleven mass spectrometers: six ion traps, two triple quadrupoles, a hybrid triple quadrupole, and two quadrupole time-of-flight instruments. Initially, 4371 spectral comparisons were carried out using the data from eleven instruments and the degree of reproducibility was evaluated. A blind trial has also been carried out to assess the reproducibility of spectra obtained during LC/MS/MS. The results suggest a degree of reproducibility across all instrument types using the tuning point technique. The reproducibility of the product ion spectra is increased when comparing the tandem in time type instruments and the tandem in space instruments as two separate groups. This may allow the production of a more limited, yet useful, screening library for LC/MS/MS identification using instruments of the same type from different manufacturers.

摘要

串联质谱实验中通过碰撞诱导解离(CID)产生的产物离子谱在不同仪器之间可能存在显著差异。已经有许多尝试来标准化产物离子谱的产生;然而,对于可重复产生谱图的最合适方法尚未达成共识。我们之前报道了在多种不同类型仪器上对产物离子谱的比较——一台三重四极杆质谱仪、两台离子阱质谱仪和一台傅里叶变换离子回旋共振质谱仪(Bristow AWT,Webb KS,Lubben AT,Halket JM. Rapid Commun. Mass Spectrom. 2004; 18: 1)。该研究表明可以实现高度的重现性。本研究的目标是提高在不同实验室使用不同仪器产生的CID产物离子质谱的可比性和重现性。这通过在每台质谱仪上为产物离子形成定义一个光谱校准点来实验性地进行。长期目标是开发一个通用的(与仪器无关的)产物离子质谱库用于未知物的鉴定。已经在11台质谱仪上记录了48种化合物的谱图:6台离子阱质谱仪、2台三重四极杆质谱仪、1台混合型三重四极杆质谱仪和2台四极杆飞行时间质谱仪。最初,使用来自11台仪器的数据进行了4371次光谱比较,并评估了重现程度。还进行了一项盲测来评估在LC/MS/MS过程中获得的谱图的重现性。结果表明使用调谐点技术在所有仪器类型中都有一定程度的重现性。当将时间串联型仪器和空间串联型仪器作为两个单独的组进行比较时,产物离子谱的重现性会提高。这可能允许使用来自不同制造商的相同类型仪器为LC/MS/MS鉴定生成一个更有限但有用的筛选库。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验