Bristow Anthony W T, Webb Kenneth S, Lubben Anneke T, Halket John
LGC Limited, Queens Road, Teddington TW11 0LY, UK.
Rapid Commun Mass Spectrom. 2004;18(13):1447-54. doi: 10.1002/rcm.1492.
The results of the comparison of product-ion tandem mass (MS/MS) spectra recorded on three ion trap mass spectrometers, a triple quadrupole mass spectrometer and a Fourier transform ion cyclotron resonance mass spectrometer are reported. The spectra were recorded in accordance with a simple experimental protocol, which involved the collision-induced dissociation (CID) attenuation of the abundance of the [M+H]+ ion to between 10 and 50% of its original abundance. The degree of similarity between the spectra from four of the mass spectrometers was calculated off-line by comparing the five most abundant ions from the spectrum on each instrument. A percentage fit value (20% for each ion that matches) was calculated by comparing each spectrum against the spectra recorded for the same compound on each instrument. The percentage of the inter-library pairwise comparisons (total = 434) that matched to > or = 60% ranged from 64-89%, depending on the instrument pair. A blind trial was also undertaken using five unknown compounds resulting in 1670 pairwise comparisons with the library entries. The blind trial produced no false positives and correct identifications in all cases. The results of the study have established the basis for the construction of a transferable product-ion MS/MS library.
报告了在三台离子阱质谱仪、一台三重四极杆质谱仪和一台傅里叶变换离子回旋共振质谱仪上记录的产物离子串联质谱(MS/MS)谱的比较结果。这些谱是按照一个简单的实验方案记录的,该方案涉及将[M+H]+离子的丰度通过碰撞诱导解离(CID)衰减至其原始丰度的10%至50%之间。通过比较每台仪器谱图中五个最丰富的离子,离线计算了四台质谱仪谱图之间的相似程度。通过将每个谱图与同一化合物在每台仪器上记录的谱图进行比较,计算出匹配度百分比值(每个匹配离子为20%)。库间成对比较(总数 = 434)中匹配度≥60%的百分比范围为64% - 89%,具体取决于仪器对。还使用五种未知化合物进行了盲测,与库条目进行了1670次成对比较。盲测未产生假阳性结果,且在所有情况下均能正确鉴定。该研究结果为构建可转移的产物离子MS/MS库奠定了基础。