Fletcher Claire Lesley, Mills John A
Department of Clinical Physics and Bioengineering, University Hospital, Coventry, CV2 2DX, UK.
Phys Med Biol. 2008 Jun 7;53(11):N209-18. doi: 10.1088/0031-9155/53/11/N02. Epub 2008 May 19.
Percentage depth dose (PDD) curves were obtained for 50 kV and 100 kV x-rays on a Gulmay Medical D3000 DXR unit. Different dosimetry systems were compared including a Scanditronix Wellhofer small volume cylindrical ion chamber, a Wellhofer photon PFD diode, a PTW soft x-ray parallel plate chamber (N23342) and two types of radiochromic film: GafChromic EBT and GafChromic MD55. The PDD curves were also compared to BEAMnrc Monte Carlo predictions. GafChromic film was found to be a valid choice of dosimeter for measuring percentage depth dose curves at 100 kV and 50 kV. All the dosimeters showed agreement with predictions at depths greater than 10 mm, while near the surface GafChromic film and PFD diodes give the best agreement to Monte Carlo values.