The longitudinal dose profile in a computed tomography dose index (CTDI) phantom had been studied by many researchers. The cross-sectional dose profile in the CTDI phantom, however, has not been studied. It is also important to understand the cross-sectional dose profile in the CTDI phantom for dose estimation in X-ray CT. In this study, the cross-sectional dose profile in the CTDI phantom was calculated by use of a Monte Carlo (MC) simulation method. A helical or a 320-detector-row cone-beam X-ray CT scanner was simulated. The cross-sectional dose profile in the CTDI phantom from surface to surface through the center point was calculated by MC simulation. The shape of the calculation region was a cylinder of 1-mm-diameter. The length of the cylinder was 23, 100, or 300 mm to represent various CT ionization chamber lengths. Detailed analyses of the energy depositions demonstrated that the cross-sectional dose profile was different in measurement methods and phantom sizes. In this study, we also focused on the validation of the weighting factor used in weighted CTDI (CTDI w ). As it stands now, the weighting factor used in CTDI w is (1/3, 2/3) for the (central, peripheral) axes. Our results showed that an equal weighting factor, which is (1/2, 1/2) for the (central, peripheral) axes, is more suitable to estimate the average cross-sectional dose when X-ray CT dose estimation is performed.
许多研究人员对计算机断层扫描剂量指数(CTDI)体模中的纵向剂量分布进行了研究。然而,CTDI体模中的横断面剂量分布尚未得到研究。了解CTDI体模中的横断面剂量分布对于X射线CT中的剂量估计也很重要。在本研究中,采用蒙特卡罗(MC)模拟方法计算了CTDI体模中的横断面剂量分布。模拟了螺旋或320排探测器的锥形束X射线CT扫描仪。通过MC模拟计算了CTDI体模中从表面到表面穿过中心点的横断面剂量分布。计算区域的形状是一个直径为1毫米的圆柱体。圆柱体的长度为23、100或300毫米,以代表不同的CT电离室长度。对能量沉积的详细分析表明,测量方法和体模尺寸不同时,横断面剂量分布也不同。在本研究中,我们还重点验证了加权CTDI(CTDI w )中使用的加权因子。目前,CTDI w 中使用的加权因子对于(中心、周边)轴为(1/3,2/3)。我们的结果表明,当进行X射线CT剂量估计时,对于(中心、周边)轴采用(1/2,1/2)的相等加权因子更适合估计平均横断面剂量。