介入放射学中主动式个人剂量计的相互比较。

Intercomparison of active personal dosemeters in interventional radiology.

作者信息

Clairand Isabelle, Struelens Lara, Bordy Jean-Marc, Daures Josiane, Debroas Jacques, Denozières Marc, Donadille Laurent, Gouriou Jean, Itié Christian, Vaz Pedro, d'Errico Francesco

机构信息

Institut de Radioprotection et de Sûreté Nucléaire (IRSN), BP17, 92262 Fontenay-aux-Roses, France.

出版信息

Radiat Prot Dosimetry. 2008;129(1-3):340-5. doi: 10.1093/rpd/ncn083. Epub 2008 May 24.

Abstract

The use of active personal dosemeters (APD) in interventional radiology was evaluated by Working Group 9 (Radiation protection dosimetry of medical staff) of the CONRAD project, which is a Coordination Action supported by the European Commission within its sixth Framework Programme. Interventional radiology procedures can be very complex and they can lead to relatively high doses to personnel who stand close to the primary radiation field and are mostly exposed to radiation scattered by the patient. For the adequate dosimetry of the scattered photons, APDs must be able to respond to low-energy [10-100 keV] and pulsed radiation with relatively high instantaneous dose rates. An intercomparison of five APD models deemed suitable for application in interventional radiology was organised in March 2007. The intercomparison used pulsed and continuous radiation beams, at CEA-LIST (Saclay, France) and IRSN (Fontenay-aux-Roses, France), respectively. A specific configuration, close to the clinical practice, was considered. The reference dose, in terms of Hp(10), was derived from air kerma measurements and from the measured and calculated energy distributions of the scattered radiation field. Additional Monte Carlo calculations were performed to investigate the energy spectra for different experimental conditions of the intercomparison. The results of this intercomparison are presented in this work and indicate which APDs are able to provide a correct response when used in the specific low-energy spectra and dose rates of pulsed X-rays encountered in interventional radiology.

摘要

CONRAD项目的第9工作组(医务人员的辐射防护剂量测定)对介入放射学中主动式个人剂量计(APD)的使用进行了评估。CONRAD项目是由欧盟委员会在其第六框架计划下支持的一项协调行动。介入放射学程序可能非常复杂,对于站在主辐射场附近且主要受到患者散射辐射照射的人员来说,可能会导致相对较高的剂量。为了对散射光子进行充分的剂量测定,APD必须能够对低能[10 - 100 keV]和具有相对较高瞬时剂量率的脉冲辐射做出响应。2007年3月组织了一次对五种被认为适用于介入放射学的APD模型的比对。该比对分别在法国CEA - LIST(萨克雷)和法国IRSN(丰特奈 - 奥罗斯)使用了脉冲和连续辐射束。考虑了一种接近临床实践的特定配置。以Hp(10)表示的参考剂量源自空气比释动能测量以及散射辐射场的测量和计算能量分布。还进行了额外的蒙特卡罗计算,以研究比对中不同实验条件下的能谱。这项工作展示了该比对的结果,并指出了哪些APD在介入放射学中遇到的特定低能谱和脉冲X射线剂量率下能够提供正确的响应。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验