全髋关节置换术中数字模板制作中X线标记物的准确性及应用评估。

Evaluation of the accuracy and use of x-ray markers in digital templating for total hip arthroplasty.

作者信息

Bayne Christopher O, Krosin Michael, Barber Thomas C

机构信息

Harvard Medical School, Boston, Massachusetts 02115, USA.

出版信息

J Arthroplasty. 2009 Apr;24(3):407-13. doi: 10.1016/j.arth.2007.11.020. Epub 2008 Apr 18.

Abstract

Despite benefits of digital templating, digital image acquisition techniques often make image magnification unknown and variable. We assessed the accuracy of digital x-ray scaling devices based on position. Radiographs of 106 total hip arthroplasty (THA) patients were studied to determine the accuracy of 2 digital x-ray markers at 1 of 4 positions by comparing true to expected x-ray magnification. The Pearson r coefficient was calculated as an indicator of measurement correlation between investigators. Greater accuracy was obtained for markers placed at the greater trochanter without skin overlap than for markers placed with soft tissue or bone overlap on x-ray. Flexible positioning markers provided greater accuracy than adhesive markers. Based on our data, we suggest positioning at the greater trochanter without skin overlap on x-ray and suggest flexible positioning over adhesive marker use.

摘要

尽管数字模板有诸多优点,但数字图像采集技术常常使图像放大倍数未知且可变。我们基于位置评估了数字X射线缩放设备的准确性。通过比较实际与预期的X射线放大倍数,研究了106例全髋关节置换术(THA)患者的X线片,以确定4个位置之一处2种数字X射线标记物的准确性。计算皮尔逊r系数作为研究者之间测量相关性的指标。与X射线上有软组织或骨骼重叠的标记物相比,放置在大转子处且无皮肤重叠的标记物具有更高的准确性。柔性定位标记物比粘贴式标记物具有更高的准确性。基于我们的数据,我们建议在X射线上定位在大转子处且无皮肤重叠,并建议使用柔性定位而非粘贴式标记物。