Prolonged inflation of coronary angioplasty balloon as treatment for subocclusive dissection of the coronary sinus during implantation of a coronary sinus pacing lead.

作者信息

Bosa Francisco, Bethencourt Miguel, Vargas Manuel, Ferrer Julio, Rodriguez Anibal, Marrero Francisco

机构信息

Cardiology, University Hospital of Canary Islands, C/ San Pedro Alcantara N degrees 20 1 degrees Iz., S/C de Tenerife, 38003, Spain.

出版信息

J Interv Card Electrophysiol. 2008 Nov;23(2):139-41. doi: 10.1007/s10840-008-9267-5. Epub 2008 Jun 6.

Abstract

Cardiac resynchronization therapy with biventricular stimulation is an accepted treatment procedure for patients with severe heart failure (NYHA class III/IV), low left ventricle ejection fraction, and left bundle branch block, in spite of optimal medical treatment (Smith, S. C. Jr, Feldman, T. E., Hirshfeld, J. W. Jr, Jacobs, A. K., Kern, M. J., King, S. B. et al. Circulation, 113(7), e166-e286, (2006)). In the ventricular resynchronization procedure, an electrode must be implanted, via the coronary sinus (CS), in a coronary vein of the left ventricle (LV). Insertion of guide catheters and guide wires through the CS towards the target vein may produce more or less extensive dissections and staining which makes visualizing the target vein opening difficult and gives rise to deferment of the procedure. We describe a case of CS dissection, produced by the catheter guide, resulting in a venous wall flap which impeded further advancement of the lead. This was treated with prolonged inflation of a coronary angioplasty balloon, which allowed completion of the procedure in the same surgical intervention. To our knowledge, this is the first report of the successful use of a coronary angioplasty balloon to resolve this complication of electrode implantation.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验