An improved impact technique for monitoring percutaneous implant integrity.

作者信息

Swain Ryan, Faulkner Gary, Raboud Don, Wolfaardt John

机构信息

Department of Mechanical Engineering, University of Alberta, Edmonton, Alberta, Canada.

出版信息

Int J Oral Maxillofac Implants. 2008 Mar-Apr;23(2):263-9.

Abstract

PURPOSE

The purpose of this study was to investigate the validity of the current Periotest system when measuring implant systems and to present a new system to monitor implant interface integrity.

MATERIALS AND METHODS

The new system records an impact accelerometer signal and utilizes software for data analysis to determine the resonance frequency of an implant-abutment system. The new system uses the handpiece from the Periotest to acquire an impact signal but makes no use of the rest of the device. Tests were completed to determine the repeatability of the new system along with the effects clinical variables such as abutment torque, angulation of the handpiece, striking height, and distance handpiece is held from the abutment have on the measurement results. Accuracy of the current Periotest method as well as the new system was independently evaluated through the use of an abutment with a strain gauge attached.

RESULTS

The new system for impact testing is shown to have greater accuracy than that of the Periotest device. Additionally, the effects of handpiece distance from abutment and torque (when above 15 Ncm) were found to be negligible while angulation of the handpiece and striking height affected the resonance frequency of the new system.

CONCLUSION

The results of the in vitro testing indicate that greater resolution and accuracy can be achieved from an impact test that utilizes a clinical measurement protocol and independent analysis of the impact accelerometer signal.

摘要

目的

本研究旨在探讨当前的牙周测试系统在测量种植体系统时的有效性,并提出一种监测种植体界面完整性的新系统。

材料与方法

新系统记录冲击加速度计信号,并利用软件进行数据分析以确定种植体-基台系统的共振频率。新系统使用牙周测试仪的手机获取冲击信号,但不使用该设备的其他部分。完成了测试以确定新系统的可重复性,以及诸如基台扭矩、手机角度、撞击高度和手机与基台之间的握持距离等临床变量对测量结果的影响。通过使用附着应变片的基台,独立评估了当前牙周测试方法以及新系统的准确性。

结果

新的冲击测试系统显示出比牙周测试设备更高的准确性。此外,发现手机与基台的距离和扭矩(当高于15 Ncm时)的影响可忽略不计,而手机角度和撞击高度会影响新系统的共振频率。

结论

体外测试结果表明,利用临床测量方案和对冲击加速度计信号进行独立分析的冲击测试可以实现更高的分辨率和准确性。