Dayan-Kenigsberg Jacqueline, Bertocchi Agnès, Garber Eric A E
Unité Biotechnologie, Biochimie des Protéines et Macromolécules, Agence Française de Sécurité Sanitaire des Produits de Santé, 143 boulevard Anatole France, 93285 Saint Denis Cedex, France.
J Immunol Methods. 2008 Jul 31;336(2):251-4. doi: 10.1016/j.jim.2008.05.007. Epub 2008 Jun 5.
Ricin can be detected in cosmetics at 0.005 microg/mL in the analytical sample using lateral flow devices (LFDs). Wheat germ, an ingredient used in skin care products is also a potential source of wheat lectin. False positives were observed when wheat lectin was added to LFDs from two manufacturers, irrespective of whether the LFD was specific for ricin, Staphylococcus enterotoxin B (SEB), or botulinum toxin. In contrast, pea and peanut lectins did not cause false positives. Substitution of the buffer supplied with the LFDs with a buffer containing 2.5% non-fat milk powder eliminated the occurrence of false positives. This substitution increased the LOD to 0.01 microg/mL ricin, which is an acceptable level for screening cosmetics for contamination by ricin.
使用侧向流动装置(LFD),在分析样品中,蓖麻毒素在化妆品中的检测限为0.005微克/毫升。小麦胚芽是护肤品中的一种成分,也是小麦凝集素的潜在来源。当从小麦凝集素添加到来自两个制造商的LFD中时,无论LFD是针对蓖麻毒素、肠毒素B(SEB)还是肉毒杆菌毒素,均观察到假阳性。相比之下,豌豆和花生凝集素不会引起假阳性。用含有2.5%脱脂奶粉的缓冲液替代LFD附带的缓冲液,可消除假阳性的出现。这种替代将蓖麻毒素的检测限提高到0.01微克/毫升,这是用于筛查化妆品中蓖麻毒素污染的可接受水平。