van Gammeren Adriaan J, van Gool Nelly, de Groot Monique J M, Cobbaert Christa M
Department of Clinical Chemistry and Hematology, Amphia Hospital, Breda, The Netherlands.
Clin Chem Lab Med. 2008;46(6):863-71. doi: 10.1515/CCLM.2008.159.
Consolidation of analyzers is an emerging issue in clinical chemistry. We evaluated the analytical performance of the Cobas 6,000 analyzer (Roche Diagnostics), which is considered a candidate for replacement of current Hitachi 917 analyzers and for consolidation of chemistry and immunochemistry.
The precision, accuracy, linearity and correlation with current field methods were evaluated according to Clinical and Laboratory Standards Institute protocols EP5, EP9 and EP10. A total of 31 routine chemistry assays and 18 immunoassays were studied. Accuracy and linearity were verified for 24 chemistry parameters using value-assigned trueness controls from the Dutch External Quality Assessment Scheme organizers. In addition, traceability to methods endorsed by the Joint Committee of Traceability in Laboratory Medicine was examined.
All analytes met allowable precision criteria, apart from the low level for sodium and folate. Total coefficients of variation ranged between 0.6% and 4.4% for routine chemistry and between 0.8% and 5.8% for immunochemistry, apart from folate (12% at the low end). The correlation coefficients for comparison to current field methods were >0.975, except for magnesium and for six out of 18 immunochemistries. Recovery experiments indicated high recovery for most of the 24 routine chemistry assays.
Considering the excellent precision data and the result equivalence for most assays, it can be concluded that Cobas 6,000 accommodates robust chemistry and immunochemistry, and has good potential for workstation consolidation in medium-sized laboratories.
分析仪的整合是临床化学领域中一个新出现的问题。我们评估了 Cobas 6000 分析仪(罗氏诊断公司)的分析性能,该分析仪被视为有望取代现有的日立 917 分析仪,并实现化学分析和免疫化学分析整合的设备。
根据临床和实验室标准协会的 EP5、EP9 和 EP10 方案,对该分析仪的精密度、准确度、线性以及与当前现场方法的相关性进行了评估。共研究了 31 项常规化学分析和 18 项免疫分析。使用荷兰外部质量评估计划组织者提供的定值真实性对照品,对 24 个化学参数的准确度和线性进行了验证。此外,还检查了其与实验室医学溯源联合委员会认可方法的溯源性。
除钠和叶酸的低水平检测外,所有分析物均符合允许的精密度标准。常规化学分析的总变异系数在 0.6%至 4.4%之间,免疫化学分析的总变异系数在 0.8%至 5.8%之间,叶酸(低端为 12%)除外。与当前现场方法相比,相关系数>0.975,但镁以及 18 项免疫分析中的 6 项除外。回收率实验表明,24 项常规化学分析中的大多数回收率较高。
考虑到出色的精密度数据以及大多数分析的结果等效性,可以得出结论,Cobas 6000 具备强大的化学分析和免疫化学分析能力,在中型实验室中具有良好的工作站整合潜力。