Hubl Walter, Zogbaum Martina, Boyd James C, Savory John, Schubert Mary, Meyer Dirk, Demant Thomas
Institut für Klinische Chemie und Laboratoriumsmedizin, Krankenhaus Dresden-Friedrichstadt, 41, D-01067 Dresden, Germany.
Clin Chim Acta. 2005 Jul 1;357(1):43-54. doi: 10.1016/j.cccn.2005.01.032. Epub 2005 Apr 15.
The Architect ci8200 is an integrated serum analyzer for photometric, electrochemical and immunological assays. Several assays of each category and the workflow performance of the system were compared with established laboratory procedures in two laboratories.
Measurements were compared with the ELECSYS 2010 (Roche Diagnostics) for CEA, PSA, FPSA, AFP, folate, vitamin B12, with the CENTAUR (Bayer) for TSH, T4, FT4, FSH and Estradiol, with the LIAISON (DiaSorin) for TSH, FT4 and FT3, with the Behring Nephelometer BN II (Dade-Behring) for ferritin, and with the INTEGRA 800 (Roche Diagnostics), and the AU640 (Olympus) for clinical chemistry assays. Workflow studies were performed to compare times of analysis required for defined analytical workloads.
The coefficients of variation (CVs) for within-run imprecision were between 3% and 6% for CEA, PSA, FPSA, AFP and ferritin, and between 3% and 11% for TSH, FT4, FT3, folate and vitamin B12. The CVs for day-to-day imprecision for immunoassays were between 3% and 10%, except for vitamin B12 (CVs 11-13%) and FT4 (CV 10% -13%). For clinical chemistry tests corresponding CVs for within-run imprecision were < 1%, except for HDL, triglyceride, creatinine, ALT, LD and lipase (CVs<2%) and bicarbonate (CV 3%-6%) and magnesium (CV < 3%). The CVs for day-to-day imprecision for clinical chemistry tests were < 1%, except for sodium, CO(2), magnesium, phosphorus, glucose, uric acid, HDL, triglyceride, ALT, AST CK, lipase with CVs < 6% and for CO(2)<11%. Dilutional linearity testing of seven immunoassays and five clinical chemistry analytes resulted in recovery rates of 90-110%. Correlation studies with 15 immunoassays and 25 clinical chemistry tests showed acceptable agreements with established methods. Work flow analyses demonstrated a net gain in time of analysis up to 109 min depending on the size of the sample batch analyzed with the Architect ci8200 as the main analyzer as compared to the currently installed routine laboratory equipment. Median turn-around times were 7 and 30 min for chemistry assays and immunoassays, respectively, when ordered as STAT analyses, and 18 min when chemistry assays were ordered as routine determinations.
Assays on the Architect ci8200 performed well, fulfilling quality control requirements as defined for instance by German quality control guidelines (RiliBAK). Method comparisons showed acceptable agreements with established assays. Workflow studies using the Architect ci8200 documented shorter times of analyses as compared with the conventionally established laboratory routine demonstrating the potential of integrated chemistry/immunoassay analyzers to provide faster and more efficient performance.
Architect ci8200是一款用于光度法、电化学法和免疫分析的集成式血清分析仪。在两个实验室中,将该系统各类别的多项检测方法及其工作流程性能与既定的实验室程序进行了比较。
将癌胚抗原(CEA)、前列腺特异抗原(PSA)、游离前列腺特异抗原(FPSA)、甲胎蛋白(AFP)、叶酸、维生素B12的检测结果与罗氏诊断公司的ELECSYS 2010进行比较,促甲状腺激素(TSH)、甲状腺素(T4)、游离甲状腺素(FT4)、促卵泡激素(FSH)和雌二醇的检测结果与拜耳公司的CENTAUR进行比较,TSH、FT4和游离三碘甲状腺原氨酸(FT3)的检测结果与索灵公司的LIAISON进行比较,铁蛋白的检测结果与达德-拜耳公司的贝林散射比浊仪BN II进行比较,临床化学检测结果与罗氏诊断公司的INTEGRA 800以及奥林巴斯公司的AU640进行比较。进行了工作流程研究,以比较规定分析工作量所需的分析时间。
CEA、PSA、FPSA、AFP和铁蛋白批内不精密度的变异系数(CV)在3%至6%之间,TSH、FT4、FT3、叶酸和维生素B12的变异系数在3%至11%之间。免疫分析的日间不精密度变异系数在3%至10%之间,维生素B12(变异系数11%-13%)和FT4(变异系数10%-13%)除外。临床化学检测的批内不精密度变异系数<1%,高密度脂蛋白(HDL)、甘油三酯、肌酐、谷丙转氨酶(ALT)、乳酸脱氢酶(LD)和脂肪酶除外(变异系数<2%),碳酸氢盐(变异系数3%-6%)和镁(变异系数<3%)。临床化学检测的日间不精密度变异系数<1%,钠、二氧化碳(CO₂)、镁、磷、葡萄糖、尿酸、HDL、甘油三酯、ALT、谷草转氨酶(AST)、肌酸激酶(CK)、脂肪酶除外,其变异系数<6%,CO₂除外(变异系数<11%)。对7项免疫分析和5种临床化学分析物进行的稀释线性测试得出的回收率为90%-110%。与15项免疫分析和25项临床化学检测的相关性研究表明,与既定方法的一致性良好。工作流程分析表明,以Architect ci8200作为主要分析仪时,与目前安装的常规实验室设备相比,根据分析样本批次的大小,分析时间净增益可达109分钟。当作为急诊分析下达指令时,化学检测和免疫分析的中位周转时间分别为7分钟和30分钟,当化学检测作为常规测定下达指令时,中位周转时间为18分钟。
Architect ci8200上的检测表现良好,符合例如德国质量控制指南(RiliBAK)所定义的质量控制要求。方法比较表明与既定检测方法的一致性良好。使用Architect ci8200进行的工作流程研究表明,与传统的实验室常规方法相比,分析时间更短,这证明了集成式化学/免疫分析仪器具有提供更快、更高效性能的潜力。