通过活动记录仪检测周期性肢体运动与多导睡眠图的比较:两种活动记录仪的验证与比较

PLM detection by actigraphy compared to polysomnography: a validation and comparison of two actigraphs.

作者信息

Gschliesser Viola, Frauscher Birgit, Brandauer Elisabeth, Kohnen Ralf, Ulmer Hanno, Poewe Werner, Högl Birgit

机构信息

Department of Neurology, Medical University of Innsbruck, Anichstrasse 35, 6020 Innsbruck, Austria.

出版信息

Sleep Med. 2009 Mar;10(3):306-11. doi: 10.1016/j.sleep.2008.03.015. Epub 2008 Jul 24.

Abstract

OBJECTIVE

To compare periodic leg movement (PLM) counts obtained with polysomnography (PSG) to those obtained from actigraphy with two devices (Actiwatch and PAM-RL).

METHODS

Twenty-four patients underwent full night actigraphy with Actiwatch from both legs and simultaneous PSG. Out of these patients, 10 had additional actigraphy with PAM-RL. Bilateral and unilateral PLM indices (PLMI) for both actigraphs were calculated for time in bed and compared to polysomnographic PLMI. Additionally, a comparison between the two different actigraphs was performed.

RESULTS

PLMI obtained with Actiwatch were significantly lower than those obtained with PSG (21.2+/-25.6/h versus 34.4+/-30.7/h; p<0.001), whereas the PLMI from PAM-RL were significantly higher than in PSG (63.6+/-39.3/h versus 37.0+/-33.5/h; p=0.009). In direct comparison, Actiwatch gave significantly lower PLMI than the PAM-RL (p=0.005). The correlations between Actiwatch and PSG (rho=0.835, p<0.001), PAM-RL and PSG (rho=0.939, p<0.001), and Actiwatch and PAM-RL (rho=0.915, p<0.001) were significant. Unilateral actigraphy compared to standard PSG gave less consistent findings. When comparing different settings of the PAM-RL, manual threshold setting resulted in PLMI that were no longer different from PSG (p=0.074), in contrast to the default threshold setting.

CONCLUSIONS

The Actiwatch underestimated and the PAM-RL overestimated PLMI compared to PSG. Whereas PLMI obtained with two actigraphs and PSG were highly correlated, they differed in mean values. Therefore, PSG, actigraphy and also the different actigraphs cannot be interchanged in longitudinal studies, and actigraphy should not be used for diagnostic decision making based on PLM indices. The best approximation to PSG PLMI was achieved by using manual threshold setting with the PAM-RL.

摘要

目的

比较通过多导睡眠图(PSG)获得的周期性腿部运动(PLM)计数与使用两种设备(Actiwatch和PAM-RL)通过活动记录仪获得的PLM计数。

方法

24例患者使用Actiwatch对双腿进行全夜活动记录,并同时进行PSG监测。其中10例患者还使用PAM-RL进行了活动记录。计算两种活动记录仪在卧床时间的双侧和单侧PLM指数(PLMI),并与多导睡眠图的PLMI进行比较。此外,还对两种不同的活动记录仪进行了比较。

结果

Actiwatch获得的PLMI显著低于PSG获得的PLMI(21.2±25.6/小时对34.4±30.7/小时;p<0.001),而PAM-RL的PLMI显著高于PSG(63.6±39.3/小时对37.0±33.5/小时;p=0.009)。直接比较时,Actiwatch的PLMI显著低于PAM-RL(p=0.005)。Actiwatch与PSG之间(rho=0.835,p<0.001)、PAM-RL与PSG之间(rho=0.939,p<0.001)以及Actiwatch与PAM-RL之间(rho=0.915,p<0.001)的相关性均显著。与标准PSG相比,单侧活动记录的结果不太一致。在比较PAM-RL的不同设置时,与默认阈值设置相比,手动阈值设置导致的PLMI与PSG不再有差异(p=0.074)。

结论

与PSG相比,Actiwatch低估了PLMI,而PAM-RL高估了PLMI。虽然两种活动记录仪和PSG获得的PLMI高度相关,但它们的平均值不同。因此,在纵向研究中,PSG、活动记录以及不同的活动记录仪不能相互替代,并且活动记录不应基于PLM指数用于诊断决策。使用PAM-RL的手动阈值设置可实现与PSG PLMI的最佳近似。