Suppr超能文献

一种用于测量睡眠中周期性腿部运动的新型活动记录仪系统的验证。

The validation of a new actigraphy system for the measurement of periodic leg movements in sleep.

作者信息

King Martin A, Jaffre Marc-Olivier, Morrish Emma, Shneerson John M, Smith Ian E

机构信息

Respiratory Support and Sleep Centre, Papworth Hospital, Papworth Everard, Cambridge CB3 8RE, UK.

出版信息

Sleep Med. 2005 Nov;6(6):507-13. doi: 10.1016/j.sleep.2004.12.010. Epub 2005 Mar 31.

Abstract

BACKGROUND AND PURPOSE

To determine the accuracy of a new actigraphy system for the measurement of periodic leg movements (PLMs). Continuous measurements of leg movements, made overnight, are essential to diagnose and monitor treatment for PLMs. We have developed the Actiwatch with Cambridge Neuro-Technology Ltd (Cambridge, UK) to detect leg movements consistent with PLMs, to record these movements from both feet over three consecutive nights and to report standard indices of PLMs. We describe three studies designed to validate this device.

PATIENTS AND METHODS

The Actiwatch was assessed on a bench model across a range of movement amplitudes. The level of agreement on individual movements between the Actiwatch and bilateral anterior tibialis electromyography (BATEMG) measures was assessed in 199 epochs from five patients with known PLMs. The ability of the Actiwatch to correctly identify patients with PLMs was assessed in a comparison with polysomnography (PSG) in 50 consecutive patients investigated in the sleep laboratory.

RESULTS

The Actiwatch detected all mechanically generated movements (100% sensitivity and 100% specificity) on the bench test. On individual movements BATEMG measurement and the Actiwatch agreed on 94% of epochs. The sensitivity and specificity of the Actiwatch to detect leg movements in severe PLMs as defined by BATEMG (PLMI>or=25) are high (100% sensitive and 97% specific). Our bilateral system gave better agreement with BATEMG than previous systems but the two measures are still not interchangeable.

CONCLUSIONS

EMG is only a surrogate measure of limb movement and indices measured in this way correlate poorly with symptoms. We have demonstrated that the Actiwatch faithfully records movement, which offers the potential to reassess the relationships between limb movement indices and symptoms. This which offers a convenient and economical alternative to PSG in the study of large populations to increase our understanding of the epidemiology and clinical significance of PLMs.

摘要

背景与目的

确定一种用于测量周期性腿部运动(PLMs)的新型活动记录仪系统的准确性。夜间对腿部运动进行连续测量对于PLMs的诊断和治疗监测至关重要。我们与英国剑桥的剑桥神经技术有限公司共同开发了活动记录仪,以检测与PLMs一致的腿部运动,连续三晚记录双脚的这些运动,并报告PLMs的标准指标。我们描述了三项旨在验证该设备的研究。

患者与方法

在一个台架模型上,对活动记录仪在一系列运动幅度下进行评估。在5名已知患有PLMs的患者的199个时段中,评估活动记录仪与双侧胫前肌肌电图(BATEMG)测量在个体运动上的一致性水平。在睡眠实验室对50名连续接受调查的患者中,将活动记录仪与多导睡眠图(PSG)进行比较,评估其正确识别PLMs患者的能力。

结果

在台架测试中,活动记录仪检测到了所有机械产生的运动(灵敏度100%,特异性100%)。在个体运动方面,BATEMG测量与活动记录仪在94%的时段上达成一致。活动记录仪检测BATEMG定义的严重PLMs(PLMI≥25)中腿部运动的灵敏度和特异性很高(灵敏度100%,特异性97%)。我们的双侧系统与BATEMG的一致性比以前的系统更好,但这两种测量方法仍不可互换。

结论

肌电图只是肢体运动的替代测量方法,以这种方式测量的指标与症状的相关性较差。我们已经证明活动记录仪能如实记录运动,这为重新评估肢体运动指标与症状之间的关系提供了可能。这为在研究大量人群时增加我们对PLMs的流行病学和临床意义的理解提供了一种方便且经济的替代PSG的方法。