Ebeid Makram R, Mehta Inder, Gaymes Charles H
Division of Cardiology, Department of Pediatrics, The University of Mississippi Medical Center, 2500 N. State Street, Jackson, MS 39216, USA.
Pediatr Cardiol. 2009 Jan;30(1):15-9. doi: 10.1007/s00246-008-9268-2. Epub 2008 Aug 19.
Newer modifications of the Fontan operation include the external tunnel conduit with placement of an elongated type of fenestration. Atrial septal defect closure devices used traditionally to close fenestrations have short connecting waists with retention discs on each side. These may not be suitable for this type of Fontan fenestration. The length of the fenestration may not allow proper expansion of the retention discs on these devices. The Amplatzer vascular plug, a new occlusion device without a centering waist or retention discs designed to close vascular malformation, can be placed entirely within an elongated tube or vessel.
A retrospective review was used to study patients who underwent catheter closure of Fontan fenestration in our institution. Only patients whose procedure was performed after the commercial release of the Amplatzer vascular plug were included in the study.
The vascular plug was implanted in 4 of 10 patients who underwent fenestration closure. The plug was successfully placed in all four patients without protrusion into either the systemic or pulmonary venous side of the baffle. There were no incidences of embolization, hemolysis, or infective endocarditis. All the patients experienced significant improvement in oxygen saturation. The technical aspects of this device are discussed as well as its safe and effective use. The satisfactory follow-up evaluation also is reported.
This study outlines a novel use of the Amplatzer vascular plug for successful closure of selected Fontan fenestrations.
Fontan手术的新型改良术式包括带有细长型开窗的体外隧道管道。传统上用于关闭开窗的房间隔缺损封堵装置连接腰部较短,两侧有固定盘。这些可能不适用于此类Fontan开窗。开窗的长度可能不允许这些装置上的固定盘适当扩张。Amplatzer血管封堵器是一种新型封堵装置,没有用于对准的腰部或固定盘,旨在关闭血管畸形,可完全放置在细长的管道或血管内。
采用回顾性研究方法,对在我院接受Fontan开窗导管封堵术的患者进行研究。本研究仅纳入在Amplatzer血管封堵器上市后接受该手术的患者。
10例接受开窗封堵术的患者中有4例植入了血管封堵器。封堵器在所有4例患者中均成功放置,未突入隔板的体静脉或肺静脉侧。没有发生栓塞、溶血或感染性心内膜炎。所有患者的血氧饱和度均有显著改善。讨论了该装置的技术细节及其安全有效的使用方法。还报告了令人满意的随访评估结果。
本研究概述了Amplatzer血管封堵器在成功关闭特定Fontan开窗方面的一种新用途。