适用于低资源心电图监测系统的在线数字滤波器和QRS波检测器。

Online digital filter and QRS detector applicable in low resource ECG monitoring systems.

作者信息

Tabakov Serafim, Iliev Ivo, Krasteva Vessela

机构信息

Department of Electronics, Technical University of Sofia, 8 Kl. Ohridski str., 1000 Sofia, Bulgaria.

出版信息

Ann Biomed Eng. 2008 Nov;36(11):1805-15. doi: 10.1007/s10439-008-9553-5. Epub 2008 Aug 28.

Abstract

The present work describes fast computation methods for real-time digital filtration and QRS detection, both applicable in autonomous personal ECG systems for long-term monitoring. Since such devices work under considerable artifacts of intensive body and electrode movements, the input filtering should provide high-quality ECG signals supporting the accurate ECG interpretation. In this respect, we propose a combined high-pass and power-line interference rejection filter, introducing the simple principle of averaging of samples with a predefined distance between them. In our implementation (sampling frequency of 250 Hz), we applied averaging over 17 samples distanced by 10 samples (Filter10x17), thus realizing a comb filter with a zero at 50 Hz and high-pass cut-off at 1.1 Hz. Filter10x17 affords very fast filtering procedure at the price of minimal computing resources. Another benefit concerns the small ECG distortions introduced by the filter, providing its powerful application in the preprocessing module of diagnostic systems analyzing the ECG morphology. Filter10x17 does not attenuate the QRS amplitude, or introduce significant ST-segment elevation/depression. The filter output produces a constant error, leading to uniform shifting of the entire P-QRS-T segment toward about 5% of the R-peak amplitude. Tests with standardized ECG signals proved that Filter10x17 is capable to remove very strong baseline wanderings, and to fully suppress 50 Hz interferences. By changing the number of the averaged samples and the distance between them, a filter design with different cut-off and zero frequency could be easily achieved. The real-time QRS detector is designed with simplified computations over single channel, low-resolution ECGs. It relies on simple evaluations of amplitudes and slopes, including history of their mean values estimated over the preceding beats, smart adjustable thresholds, as well as linear logical rules for identification of the R-peaks in real-time. The performance of the QRS detector was tested with internationally recognized ECG databases (AHA, MIT-BIH, European ST-T database), showing mean sensitivity of 99.65% and positive predictive value of 99.57%. The performance of the presented QRS detector can be highly rated, comparable and even better than other published real-time QRS detectors. Examples representing some typical unfavorable conditions in real ECGs, illustrate the common operation of Filter10x17 and the QRS detector.

摘要

本研究描述了用于实时数字滤波和QRS波检测的快速计算方法,这两种方法都适用于用于长期监测的自主式个人心电图系统。由于此类设备在身体剧烈运动和电极移动产生的大量伪影下工作,输入滤波应提供高质量的心电图信号,以支持准确的心电图解读。在这方面,我们提出了一种组合的高通和电源干扰抑制滤波器,引入了对样本进行平均的简单原理,样本之间有预定义的距离。在我们的实现中(采样频率为250Hz),我们对相隔10个样本的17个样本进行平均(Filter10x17),从而实现了一个在50Hz处有零点且高通截止频率为1.1Hz的梳状滤波器。Filter10x17以最小的计算资源为代价提供了非常快速的滤波过程。另一个好处是该滤波器引入的心电图失真很小,这使其在分析心电图形态的诊断系统的预处理模块中具有强大的应用价值。Filter10x17不会衰减QRS波幅度,也不会引入明显的ST段抬高/压低。滤波器输出会产生一个恒定误差,导致整个P-QRS-T段朝着R波峰值幅度的约5%均匀偏移。使用标准化心电图信号进行的测试证明,Filter10x17能够消除非常强烈的基线漂移,并完全抑制50Hz干扰。通过改变平均样本的数量及其之间的距离,可以轻松实现具有不同截止频率和零点频率的滤波器设计。实时QRS波检测器针对单通道、低分辨率心电图进行了简化计算设计。它依赖于对幅度和斜率的简单评估,包括对前序心搏估计的平均值历史、智能可调阈值以及用于实时识别R波峰的线性逻辑规则。使用国际认可的心电图数据库(AHA、MIT-BIH、欧洲ST-T数据库)对QRS波检测器的性能进行了测试,结果显示平均灵敏度为99.65%,阳性预测值为99.57%。所展示的QRS波检测器的性能可得到高度评价,与其他已发表的实时QRS波检测器相当,甚至更好。代表实际心电图中一些典型不利情况的示例说明了Filter10x17和QRS波检测器的常见操作。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验