Lin Jing-Fung
Department of Computer Application Engineering, Far East University, No. 49, Jhonghua Road, Sinshih Township, Tainan County 744, Taiwan.
Appl Opt. 2008 Sep 1;47(25):4529-39. doi: 10.1364/ao.47.004529.
A novel technique is presented for obtaining concurrent measurements of the linear and circular birefringence properties of an optical sample by using a rotating-wave-plate Stokes polarimeter to extract the 2x2 central elements of the corresponding Mueller matrix via two linearly polarized probe lights. For a compound sample comprising a half-wave plate in series with a quarter-wave plate, the measured values of the principal angle and retardance of the quarter-wave plate are found to have average normalized errors of 0.56% and 1.16%, respectively, while the measured value of the rotation angle of the half-wave plate has an error of just 0.39%. When analyzing glucose solutions with concentrations ranging from 0-1.2 g/dl positioned in front of a half-wave plate, the average normalized errors in the principal axis angle and retardance measurements of the half-wave plate are 0.69% and 2.65%, respectively, while the error in the rotation angle measurements of the glucose solutions is 2.13%. The correlation coefficient between the measured rotation angle and the concentration of the glucose solution is determined to be 0.99985, while the standard deviation is just 0.0022 deg. Overall the experimental results demonstrate the ability of the proposed system to obtain highly accurate measurements of the linear and circular birefringence properties of an optical sample and to decouple the relationship between the principal axis angle and the rotation angle.
提出了一种新技术,通过使用旋转波片斯托克斯偏振计,利用两束线偏振探测光提取相应穆勒矩阵的2x2中心元素,来同时测量光学样品的线性和圆双折射特性。对于由一个半波片和一个四分之一波片串联组成的复合样品,发现四分之一波片的主角度和延迟的测量值的平均归一化误差分别为0.56%和1.16%,而半波片的旋转角度测量值的误差仅为0.39%。当分析置于半波片前浓度范围为0 - 1.2 g/dl的葡萄糖溶液时,半波片的主轴角度和延迟测量的平均归一化误差分别为0.69%和2.65%,而葡萄糖溶液旋转角度测量的误差为2.13%。测得的葡萄糖溶液旋转角度与浓度之间的相关系数确定为0.99985,标准差仅为0.0022度。总体而言,实验结果证明了所提出的系统能够高精度测量光学样品的线性和圆双折射特性,并解耦主轴角度和旋转角度之间的关系。