Brock Kristy K, Hawkins Maria, Eccles Cynthia, Moseley Joanne L, Moseley Douglas J, Jaffray David A, Dawson Laura A
Radiation Medicine Program, Princess Margaret Hospital, University Health Network and University of Toronto, Toronto, Ontario, Canada.
Acta Oncol. 2008;47(7):1279-85. doi: 10.1080/02841860802256491.
To quantify the improvements in online target localization using kV cone beam CT (CBCT) with deformable registration.
Twelve patients treated under a 6 fraction liver cancer radiation therapy protocol were imaged in breath hold using kV CBCT at each treatment fraction. The images were imported into the treatment planning software and rigidly registered by fitting the liver, identified on the daily kV CBCT image, into the liver contours, previously drawn on the planning CT. The liver was then manually contoured on each CBCT image. Deformable registration was automatically performed, aligning the CT liver to the liver on each CBCT image using MORFEUS, a biomechanical model based deformable registration algorithm. The tumor, defined on planning CT, was mapped onto the CBCT, through MORFEUS. The center of mass (COM) displacement of the tumor was computed.
The mean (SD) displacement magnitude (absolute value) of the COM following deformable registration was 0.08 (0.07), 0.10 (0.11), and 0.10 (0.17) cm in the left-right (LR), anterior-posterior (AP), and superior-inferior (SI) directions, respectively. The maximum displacement of the COM was 0.34, 0.65, and 0.97 cm in the LR, AP, and SI directions, respectively. Fifteen percent of the treatment fractions had a COM displacement of greater than 0.3 cm and 33% of patients had at least 1 fraction with a displacement of greater than 0.3 cm. The deformable registration, excluding the manual contouring of the liver, was performed in less than 1 minute, on average.
Rigid registration of the liver volume between planning CT and verification kV CBCT localizes the tumor to within 0.3 cm for the majority (66%) of patients; however, larger offsets in tumor position can be observed due to liver deformation.
使用具有可变形配准功能的千伏锥形束CT(CBCT)对在线靶区定位的改善情况进行量化。
12例按照6分割肝癌放射治疗方案接受治疗的患者,在每个治疗分割时屏气状态下使用千伏CBCT进行成像。将图像导入治疗计划软件,并通过将每日千伏CBCT图像上识别出的肝脏与先前在计划CT上绘制的肝脏轮廓进行拟合,进行刚性配准。然后在每个CBCT图像上手动勾勒肝脏轮廓。使用MORFEUS(一种基于生物力学模型的可变形配准算法)自动进行可变形配准,将CT肝脏与每个CBCT图像上的肝脏对齐。通过MORFEUS将计划CT上定义的肿瘤映射到CBCT上。计算肿瘤的质心(COM)位移。
可变形配准后,质心在左右(LR)、前后(AP)和上下(SI)方向上的平均(标准差)位移大小(绝对值)分别为0.08(0.07)、0.10(0.11)和0.10(0.17)cm。质心在LR、AP和SI方向上的最大位移分别为0.34、0.65和0.97 cm。15%的治疗分割质心位移大于0.3 cm,33%的患者至少有1个分割的位移大于0.3 cm。平均而言,可变形配准(不包括肝脏的手动勾勒)在不到1分钟内完成。
计划CT和验证千伏CBCT之间肝脏体积的刚性配准可将大多数(66%)患者的肿瘤定位在0.3 cm范围内;然而,由于肝脏变形,可观察到肿瘤位置有较大偏移。