Characterization of flow cytometer instrument sensitivity.

作者信息

Hoffman Robert A, Wood James C S

机构信息

BD Biosciences, San Jose, California, USA.

出版信息

Curr Protoc Cytom. 2007 Apr;Chapter 1:Unit1.20. doi: 10.1002/0471142956.cy0120s40.

Abstract

Fluorescence sensitivity, measured in terms of resolution, allows the researcher to more accurately answer the question of how dim a cell can be and still be resolvable from another population. This measure focuses on the width, i.e., the standard deviation, of the population distributions and not on the location, i.e., the mean intensity, of populations on a histogram scale relative to the background. To determine the fluorescence sensitivity in terms of the resolution, both the detection efficiency, Q, and optical background, B, need to be measured. Both factors affect the ability to resolve dimly fluorescent subpopulations, and both factors are required to uniquely characterize the performance of a flow cytometer. Using Q and B, it is possible to determine the minimum fluorescence intensity that is resolvable from the background or from another population. This unit describes a practical and robust approach to measure the critical factors that determine the ability of a flow cytometer to analyze dimly fluorescent particles.

摘要

以分辨率衡量的荧光灵敏度,使研究人员能够更准确地回答这样一个问题:一个细胞能有多暗,并且仍能与其他群体区分开来。该测量关注的是群体分布的宽度,即标准差,而不是直方图尺度上相对于背景的群体位置,即平均强度。为了根据分辨率确定荧光灵敏度,需要测量检测效率Q和光学背景B。这两个因素都会影响分辨微弱荧光亚群的能力,并且这两个因素都是唯一表征流式细胞仪性能所必需的。利用Q和B,可以确定可从背景或其他群体中分辨出的最小荧光强度。本单元描述了一种实用且稳健的方法,用于测量决定流式细胞仪分析微弱荧光颗粒能力的关键因素。