Gollee H, Hunt K J, Allan D B, Fraser M H, McLean A N
Centre for Rehabilitation Engineering, University of Glasgow, UK.
Technol Health Care. 2008;16(4):273-81.
Paralysis of the respiratory muscles in people with tetraplegia affects their ability to breathe and contributes to respiratory complications. Surface functional electrical stimulation (FES) of abdominal wall muscles can be used to increase tidal volume (V_{T}) and improve cough peak flow (CPF) in tetraplegic subjects who are able to breathe spontaneously. This study aims to evaluate the feasibility and effectiveness of a novel abdominal FES system which generates stimulation automatically, synchronised with the subjects' voluntary breathing activity. Four subjects with complete tetraplegia (C4-C6), breathing spontaneously, were recruited. The automatic stimulation system ensured that consistent stimulation was achieved. We compared spirometry during unassisted and FES-assisted quiet breathing and coughing, and measured the effect of stimulation on end-tidal CO_2 (EtCO_2) during quiet breathing. The system dependably recognised spontaneous respiratory effort, stimulating appropriately, and was well tolerated by patients. Significant increases in V_T during quiet breathing (range 0.05-0.23 L) and in CPF (range 0.04-0.49 L/s) were observed. Respiratory rate during quiet breathing decreased in all subjects when stimulated, whereas minute ventilation increased by 1.05-2.07 L/min. The changes in EtCO_2 were inconclusive. The automatic stimulation system augmented spontaneous breathing and coughing in tetraplegic patients and may provide a potential means of respiratory support for tetraplegic patients with reduced respiratory capacity.
四肢瘫痪患者呼吸肌麻痹会影响其呼吸能力,并导致呼吸并发症。对于能够自主呼吸的四肢瘫痪患者,可通过对腹壁肌肉进行表面功能性电刺激(FES)来增加潮气量(VT)并改善咳嗽峰流速(CPF)。本研究旨在评估一种新型腹部FES系统的可行性和有效性,该系统能自动产生与受试者自主呼吸活动同步的刺激。招募了4名完全性四肢瘫痪(C4 - C6)且能自主呼吸的受试者。自动刺激系统确保了一致的刺激效果。我们比较了无辅助和FES辅助下安静呼吸及咳嗽时的肺功能测定结果,并测量了安静呼吸时刺激对呼气末二氧化碳(EtCO2)的影响。该系统能可靠地识别自主呼吸努力并进行适当刺激,且患者耐受性良好。观察到安静呼吸时VT显著增加(范围为0.05 - 0.23升),CPF也显著增加(范围为0.04 - 0.49升/秒)。刺激时所有受试者安静呼吸时的呼吸频率均降低,而分钟通气量增加了1.05 - 2.07升/分钟。EtCO2的变化尚无定论。自动刺激系统增强了四肢瘫痪患者的自主呼吸和咳嗽能力,可能为呼吸能力下降的四肢瘫痪患者提供一种潜在的呼吸支持手段。