An iterative reconstruction technique for geometric distortion-corrected segmented echo-planar imaging.

作者信息

Xu Yingbiao, Haacke E Mark

机构信息

Department of Electric and Computer Engineering, McMaster University, Hamilton, Ontario, Canada.

出版信息

Magn Reson Imaging. 2008 Dec;26(10):1406-14. doi: 10.1016/j.mri.2008.04.008. Epub 2008 Sep 9.

Abstract

In this article, we present a modified interleaved segmented echo-planar imaging (SEPI) sequence with a center-out k-space trajectory that is especially designed for susceptibility-weighted imaging applications. We introduce a simple and efficient technique to phase correct the acquired SEPI data in the presence of moderate field inhomogeneities. This phase correction reduces the distortion in the phase-encoding direction without requiring an extra reference scan. With the use of a center-out k-space trajectory and a low-spatial-frequency phase map, phase discontinuities between segments can be eliminated, in principle, iteratively using a fast Fourier transform from the center segment to the outermost segment in k-space. With an extra echo added in front of the echo train, neither phase unwrapping nor an extra reference scan is required to obtain a low-spatial-frequency phase map. The method is shown to remove blurring and reduce geometric distortion caused by phase changes from echo to echo in both phantom and human data. The method is most useful for high-resolution imaging applications and moderate factors of speed improvement.

摘要

相似文献

1
An iterative reconstruction technique for geometric distortion-corrected segmented echo-planar imaging.
Magn Reson Imaging. 2008 Dec;26(10):1406-14. doi: 10.1016/j.mri.2008.04.008. Epub 2008 Sep 9.
2
Application of k-space energy spectrum analysis to susceptibility field mapping and distortion correction in gradient-echo EPI.
Neuroimage. 2006 Jun;31(2):609-22. doi: 10.1016/j.neuroimage.2005.12.022. Epub 2006 Feb 15.
3
EPI image reconstruction with correction of distortion and signal losses.
J Magn Reson Imaging. 2006 Sep;24(3):683-9. doi: 10.1002/jmri.20672.
5
Pseudo multishot echo-planar imaging for geometric distortion improvement.
NMR Biomed. 2023 May;36(5):e4885. doi: 10.1002/nbm.4885. Epub 2022 Dec 16.

引用本文的文献

1
Susceptibility-weighted imaging: current status and future directions.
NMR Biomed. 2017 Apr;30(4). doi: 10.1002/nbm.3552. Epub 2016 May 18.
2
Rapid brain MRI acquisition techniques at ultra-high fields.
NMR Biomed. 2016 Sep;29(9):1198-221. doi: 10.1002/nbm.3478. Epub 2016 Feb 2.
3
Initial clinical experience with a radiation oncology dedicated open 1.0T MR-simulation.
J Appl Clin Med Phys. 2015 Mar 8;16(2):5201. doi: 10.1120/jacmp.v16i2.5201.
4
MR venography of the fetal brain using susceptibility weighted imaging.
J Magn Reson Imaging. 2014 Oct;40(4):949-57. doi: 10.1002/jmri.24476. Epub 2013 Nov 25.
5
Improved MR venography using quantitative susceptibility-weighted imaging.
J Magn Reson Imaging. 2014 Sep;40(3):698-708. doi: 10.1002/jmri.24413. Epub 2013 Oct 31.
6
High-resolution structural and functional assessments of cerebral microvasculature using 3D Gas ΔR2*-mMRA.
PLoS One. 2013 Nov 4;8(11):e78186. doi: 10.1371/journal.pone.0078186. eCollection 2013.
7
Brain iron detected by SWI high pass filtered phase calibrated with synchrotron X-ray fluorescence.
J Magn Reson Imaging. 2010 Jun;31(6):1346-54. doi: 10.1002/jmri.22201.

本文引用的文献

1
A homogeneity correction method for magnetic resonance imaging with time-varying gradients.
IEEE Trans Med Imaging. 1991;10(4):629-37. doi: 10.1109/42.108599.
2
A model-based method for phase unwrapping.
IEEE Trans Med Imaging. 1996;15(6):893-7. doi: 10.1109/42.544507.
3
Susceptibility weighted imaging (SWI).
Magn Reson Med. 2004 Sep;52(3):612-8. doi: 10.1002/mrm.20198.
4
Reducing oblique flow effects in interleaved EPI with a centric reordering technique.
Magn Reson Med. 2001 Apr;45(4):623-9. doi: 10.1002/mrm.1084.
5
Optimized distortion correction technique for echo planar imaging.
Magn Reson Med. 2001 Mar;45(3):525-8. doi: 10.1002/1522-2594(200103)45:3<525::aid-mrm1070>3.0.co;2-s.
6
Motion correction with PROPELLER MRI: application to head motion and free-breathing cardiac imaging.
Magn Reson Med. 1999 Nov;42(5):963-9. doi: 10.1002/(sici)1522-2594(199911)42:5<963::aid-mrm17>3.0.co;2-l.
7
Off-resonance correction of MR images.
IEEE Trans Med Imaging. 1999 Jun;18(6):481-95. doi: 10.1109/42.781014.
8
Correction for EPI distortions using multi-echo gradient-echo imaging.
Magn Reson Med. 1999 Jun;41(6):1206-13. doi: 10.1002/(sici)1522-2594(199906)41:6<1206::aid-mrm17>3.0.co;2-l.
9
Correction of off resonance-related distortion in echo-planar imaging using EPI-based field maps.
Magn Reson Med. 1998 Feb;39(2):328-30. doi: 10.1002/mrm.1910390223.
10
Reduction of geometric and intensity distortions in echo-planar imaging using a multireference scan.
Magn Reson Med. 1997 Jun;37(6):932-42. doi: 10.1002/mrm.1910370619.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验