J Orthop Sports Phys Ther. 1989;10(11):441-7. doi: 10.2519/jospt.1989.10.11.441.
This research was supported by a grant from The Foundation for Physical Therapy, Alexandria, VA. The purpose of this study was to determine the reliability of a quantitative measure of knee ligament laxity in vivo. The testing device simultaneously measures applied forces and resulting displacement at the knee through use of a multidimensional electrogoniometer, dynamometer, and a computer-assisted analysis system. Subjects were 30 females and 20 males, ranging in age from 18-32 years. Laxity tests were performed in the anterior/posterior (A/P) plane at 90 and 30 degrees of knee flexion, and the varus/valgus (V/V) plane at 20 and 0 degrees of knee flexion. Test-retest (N = 41) and intertester (N = 18) repeatability was assessed in all test positions by ANOVA and by coefficients of correlation and variability. No significant differences were found within testers for A/P or V/V tests 1 week apart at any angle tested. There were significant differences between testers for A/P and V/V tests. No significant differences were found between left and right sides, or between male/female values. Comparisons to other in vitro and in vivo studies are made. J Orthop Sports Phys Ther 1989;10(11):441-447.
这项研究得到了弗吉尼亚州亚历山大市物理治疗基金会的资助。本研究旨在确定一种活体测量膝关节韧带松弛的定量方法的可靠性。该测试设备通过使用多维电子测角计、测力计和计算机辅助分析系统,同时测量膝关节的施加力和产生的位移。受试者为 30 名女性和 20 名男性,年龄在 18-32 岁之间。松弛度测试在膝关节屈曲 90 度和 30 度的前/后(A/P)平面以及膝关节屈曲 20 度和 0 度的内/外翻(V/V)平面上进行。通过方差分析以及相关系数和变异性系数,评估了所有测试位置的测试者内(N = 41)和测试者间(N = 18)的可重复性。在任何测试角度下,相隔 1 周的 A/P 或 V/V 测试中,测试者之间均未发现显著差异。A/P 和 V/V 测试中,测试者之间存在显著差异。左右两侧或男女之间均无显著差异。与其他体外和体内研究进行了比较。J Orthop Sports Phys Ther 1989;10(11):441-447。