Shiraishi Steve, Lee Paul W N, Leung Andrew, Goh Victor H H, Swerdloff Ronald S, Wang Christina
General Clinical Research Center, Division of Endocrinology and Metabolism, Harbor-UCLA Medical Center and Los Angeles Biomedical Research Institute, Torrance, CA 90509, USA.
Clin Chem. 2008 Nov;54(11):1855-63. doi: 10.1373/clinchem.2008.103846. Epub 2008 Sep 18.
Recent reports have described inherent problems with androgen immunoassays compared with mass spectrometry analyses.
We developed a method for measuring serum testosterone (T) and 5alpha-dihydrotestosterone (DHT) simultaneously via liquid-liquid extraction followed by liquid chromatography-tandem mass spectrometry (LC-MS/MS) with positive-mode electrospray ionization.
The DHT and T calibrators showed a linear response from 0.069 nmol/L to 34.4 nmol/L and 69.3 nmol/L, respectively. T interference in the DHT assay and vice versa were negligible. Within- and between-run imprecision values were <5% for both analytes. Percent recoveries of T and DHT spiked into samples at concentrations spanning the calibration curve were 100%-113% and 98%-107%, respectively. The lower limit of quantification was 0.069 nmol/L for both steroids. Serum T concentrations measured by LC-MS/MS were different from those obtained by RIA, especially at lower T concentrations. Serum DHT concentrations measured by LC-MS/MS were markedly lower than those generated by RIA because of the nonselectivity of the RIA without chromatography. The reference intervals (mean +/- 2 SDs) determined for T and DHT were 9.2-33.7 nmol/L and 0.47-2.65 nmol/L, respectively, for 113 healthy adult men and 0.33-2.02 nmol/L and 0.09-0.91 nmol/L, respectively, for 133 healthy premenopausal women.
We have developed and validated a selective and precise method for simultaneous measurements of serum T and DHT that can be adopted for routine measurements of these androgens in health and disease in men and women.
近期报告指出,与质谱分析相比,雄激素免疫测定存在固有问题。
我们开发了一种方法,通过液-液萃取,然后采用正模式电喷雾电离的液相色谱-串联质谱法(LC-MS/MS)同时测量血清睾酮(T)和5α-双氢睾酮(DHT)。
DHT和T校准品分别在0.069 nmol/L至34.4 nmol/L和69.3 nmol/L范围内呈线性响应。DHT测定中T的干扰以及T测定中DHT的干扰均可忽略不计。两种分析物的批内和批间不精密度值均<5%。在校准曲线范围内的浓度下添加到样品中的T和DHT的回收率分别为100%-113%和98%-107%。两种类固醇的定量下限均为0.069 nmol/L。通过LC-MS/MS测量的血清T浓度与通过放射免疫分析(RIA)获得的浓度不同,尤其是在较低的T浓度下。由于RIA没有色谱分离的非选择性,通过LC-MS/MS测量的血清DHT浓度明显低于RIA测量的浓度。113名健康成年男性的T和DHT的参考区间(平均值±2标准差)分别为9.2-33.7 nmol/L和0.47-2.65 nmol/L,133名健康绝经前女性的参考区间分别为0.33-2.02 nmol/L和0.09-0.91 nmol/L。
我们开发并验证了一种选择性和精确的方法,可同时测量血清T和DHT,可用于男性和女性健康与疾病中这些雄激素的常规测量。