Department of Laboratory Medicine, Konventhospital Barmherzige Brueder, Linz, Austria.
Clin Appl Thromb Hemost. 2009 Dec;15(6):652-9. doi: 10.1177/1076029608322547. Epub 2008 Sep 19.
This study evaluated the utility of the PFA-100 and the Multiplate analyzer for the assessment of aspirin and clopidogrel effects on platelet function in patients with cardiovascular disease. Platelet function was determined with the PFA-100 using collagen+epinephrine (CEPI) and collagen+adenosine-5'-diphosphate (CADP) cartridges, and with whole blood impedance aggregometry using the Multiplate ASPI and ADP+PG tests (aggregation triggered with arachidonic acid and ADP+ prostaglandin E1, respectively). Four study groups were identified from the 154 patients enrolled: patients without antiplatelet therapy, patients with 100 mg aspirin daily but without clopidogrel treatment, patients with 75 mg clopidogrel daily but without aspirin treatment, and patients with both 100 mg aspirin daily plus 75 mg clopidogrel daily. It was found that the PFA-100 instrument is useful for detection of aspirin but not for detection of a clopidogrel effect, while the Multiplate analyzer is useful for specific detection of both aspirin and clopidogrel effects on platelet function.
本研究评估了 PFA-100 和 Multiplate 分析仪在评估心血管疾病患者阿司匹林和氯吡格雷对血小板功能影响方面的应用。使用 PFA-100 仪通过胶原+肾上腺素 (CEPI) 和胶原+腺苷二磷酸 (CADP) 检测试剂盒,以及通过全血阻抗聚集仪使用 Multiplate ASPI 和 ADP+PG 检测试剂盒(分别用花生四烯酸和 ADP+前列腺素 E1 触发聚集)来确定血小板功能。从纳入的 154 名患者中确定了四个研究组:无抗血小板治疗的患者、每日服用 100 毫克阿司匹林但无氯吡格雷治疗的患者、每日服用 75 毫克氯吡格雷但无阿司匹林治疗的患者,以及每日服用 100 毫克阿司匹林加 75 毫克氯吡格雷的患者。结果发现,PFA-100 仪器可用于检测阿司匹林,但不能用于检测氯吡格雷的作用,而 Multiplate 分析仪可用于特异性检测阿司匹林和氯吡格雷对血小板功能的影响。