Derosier C, Delegue G, Munier T, Pharaboz C, Cosnard G
Service de Radiologie, HIA Val-de-Grâce, Paris.
J Radiol. 1991 Jun-Jul;72(6-7):349-53.
The MRI technology may be the starting-point of geometric distorsion. The mathematical preciseness of a spatial location may be disturbed and alter the guidance of an MRI interventionnal act, especially in stereotactic brain biopsy. A review of the literature shows errors of 1 to 1.5 mm. Our results show an error of 0.16 +/- 0.66 mm. The control of quality: homogeneity and calibration of magnetic-field gradients, permit an improve of the balistic preciseness and give permission to realize the guidance of a stereotactic brain biopsy with the alone MRI.
MRI技术可能是几何畸变的起始点。空间位置的数学精确性可能受到干扰,并改变MRI介入操作的引导,尤其是在立体定向脑活检中。文献综述显示误差为1至1.5毫米。我们的结果显示误差为0.16 +/- 0.66毫米。质量控制:磁场梯度的均匀性和校准,可提高弹道精确性,并允许仅使用MRI实现立体定向脑活检的引导。