Use of the Inspiron nebuliser during continuous positive airway pressure ventilation.

作者信息

Hardwick M, Dickson G, Clutton-Brock T

机构信息

Birmingham Accident Hospital.

出版信息

Anaesthesia. 1991 Aug;46(8):664-7. doi: 10.1111/j.1365-2044.1991.tb09719.x.

Abstract

The Inspiron Nebuliser 002305 with air entrainment was assessed as a gas delivery device in a continuous positive airway pressure system. Inspired oxygen concentrations, total gas flows and pressures within the system were measured over a range of settings, with and without positive and expiratory pressure. Inspired oxygen concentrations and total flows were completely disrupted when a positive and expiratory pressure valve was applied, and the system failed to generate continuous positive airway pressure. We would not therefore recommend the Inspiron nebuliser as a gas delivery system for continuous positive airway pressure.

摘要

带有空气夹带装置的灵越雾化器002305在持续气道正压通气系统中被评估为一种气体输送设备。在一系列设置条件下,测量了系统内的吸入氧浓度、总气体流量和压力,包括有和没有正压及呼气末正压的情况。当应用呼气末正压阀时,吸入氧浓度和总流量完全紊乱,并且该系统无法产生持续气道正压。因此,我们不推荐将灵越雾化器作为持续气道正压通气的气体输送系统。