Strong E Bradley, Rafii Amir, Holhweg-Majert Bettina, Fuller Scott C, Metzger Marc Christian
Department of Otolaryngology, University of California, Davis, School of Medicine, 2521 Stockton Blvd, Ste 7200, Sacramento, CA 95817, USA.
Arch Otolaryngol Head Neck Surg. 2008 Oct;134(10):1080-4. doi: 10.1001/archotol.134.10.1080.
To compare the accuracy of 3 computer-aided surgery systems for maxillofacial reconstruction.
Evaluation of 3 computer-aided surgery systems: StealthStation, VectorVision, and Voxim.
The University of California, Davis, Department of Otolaryngology computer-aided surgery laboratory.
Four fresh cadaveric heads.
Mean target registration error.
The StealthStation was the most accurate (mean [SD] target registration error, 1.00 [0.04] mm), followed by VectorVision (1.13 [0.05] mm) and then Voxim (1.34 [0.04] mm). All values met statistical significance (P < .05).
Measurable accuracy differences were found among the navigation systems evaluated. The StealthStation was the most accurate. However, the differences are small, and the clinical significance for maxillofacial reconstruction is negligible.