La Marca Frank, Zubay Geoffrey, Morrison Thomas, Karahalios Dean
Department of Neurosurgery, University of Michigan Health System, Ann Arbor, Michigan 48109-5338, USA.
J Neurosurg Spine. 2008 Oct;9(4):347-53. doi: 10.3171/SPI.2008.9.10.347.
The occipital condyle has never been studied as a viable structure that could permit bone purchase by fixation devices for occipitocervical fusion. The authors propose occipital condyle screw placement as a possible alternative to conventional occipitocervical fixation techniques.
Six adult cadaver heads (12 total occipital condyles) were studied, and the StealthStation image-guidance system was used for preoperative planning of occipital condyle screw placement. Morphometric studies of the occipital condyle were performed. A 3.5-mm Vertex screw was then placed in the occipital condyle with image-guided assistance in 3 specimens. Operations in the remaining 3 specimens proceeded using anatomical markers and calculated degrees of angulation for screw placement with a free-hand technique. Postoperatively the cadaver heads were rescanned and reanalyzed to determine the success of screw placement and its effect on hypoglossal canal volume.
All screws were successfully placed with no sign of lateral or medial cortical breach. Two screws had bicortical purchase. There was no change in hypoglossal canal volume in any specimen.
Occipital condyle screw placement is a safe and viable option for occipitocervical fixation and could be a preferred procedure in selected cases. However, further biomechanical studies are required to compare its reliability to other more established techniques.
枕髁从未被作为一种可行的结构进行研究,这种结构可允许固定装置用于枕颈融合时获得骨质锚固。作者提出枕髁螺钉置入术可作为传统枕颈固定技术的一种可能替代方法。
对6个成人尸体头部(共12个枕髁)进行研究,使用StealthStation图像引导系统进行枕髁螺钉置入的术前规划。对枕髁进行形态学研究。然后在3个标本中借助图像引导将一枚3.5毫米的Vertex螺钉置入枕髁。其余3个标本的手术采用解剖标志并通过徒手技术计算螺钉置入的角度进行。术后对尸体头部再次进行扫描和分析,以确定螺钉置入是否成功及其对舌下神经管容积的影响。
所有螺钉均成功置入,未出现外侧或内侧皮质破裂迹象。两枚螺钉实现了双侧皮质锚固。所有标本的舌下神经管容积均无变化。
枕髁螺钉置入术是枕颈固定的一种安全可行的选择,在某些特定病例中可能是首选方法。然而,需要进一步进行生物力学研究以将其可靠性与其他更成熟的技术进行比较。