Pultar Johanna, Sauer Ursula, Domnanich Patrick, Preininger Claudia
Austrian Research Centers GmbH - ARC, Department of Bioresources, 2444 Seibersdorf, Austria.
Biosens Bioelectron. 2009 Jan 1;24(5):1456-61. doi: 10.1016/j.bios.2008.08.052. Epub 2008 Sep 11.
This study describes a RNA aptamer-based biochip with high affinity and specificity for C-reactive protein (CRP). CRP, which exists in concentrations of 1-3mg/l in the serum of healthy patients, has been identified as a reliable biomarker for inflammation and as a potential marker for sepsis and tissue necrosis. The CRP-specific aptamer was covalently immobilized with its 5'-end on ARChip Epoxy. The detection of bound CRP was carried out optically using labelled secondary antibody in a sandwich format. Assay conditions were optimized with respect to the CRP binding buffer (buffer system, pH and additives) and Ca(2+) concentration (10 mM). Moreover, two sandwich immunoassay formats were tested, the one using dye-labelled antibodies and the other with biotin-modified antibodies/Dy647-labelled streptavidin. In comparison with an antibody-based chip assay, the aptamer chip is superior in terms of CRP measuring range (10 microg/l to 100mg/l) in human serum whereas antibody-based chips result in superior data reproducibility (CV of 8-15%). In contrast to antibody chips, aptamer microarrays provide the unique potential of detecting CRP in serum samples of low risk patients (1-3mg/l) as well as high risk patients (>500 mg/l), furthermore elevated CRP levels (20-350 mg/l) with acceptable recovery (70-130%) by including only one serum sample dilution step (1:100) for the complete measuring range.
本研究描述了一种对C反应蛋白(CRP)具有高亲和力和特异性的基于RNA适体的生物芯片。CRP在健康患者血清中的浓度为1-3mg/l,已被确定为炎症的可靠生物标志物以及脓毒症和组织坏死的潜在标志物。CRP特异性适体通过其5'-末端共价固定在ARChip Epoxy上。使用标记的二抗以夹心形式通过光学方法检测结合的CRP。针对CRP结合缓冲液(缓冲液系统、pH值和添加剂)和Ca(2+)浓度(10 mM)对测定条件进行了优化。此外,测试了两种夹心免疫测定形式,一种使用染料标记的抗体,另一种使用生物素修饰的抗体/Dy647标记的链霉亲和素。与基于抗体的芯片测定相比,适体芯片在人血清中CRP测量范围(10微克/升至100毫克/升)方面更具优势,而基于抗体的芯片则具有更高的数据重现性(CV为8-15%)。与抗体芯片不同,适体微阵列具有独特的潜力,能够检测低风险患者(1-3mg/l)以及高风险患者(>500mg/l)血清样本中的CRP,此外,通过仅对整个测量范围进行一步血清样本稀释(1:100),即可检测到升高的CRP水平(20-350mg/l),回收率可接受(70-130%)。