Reise Katie, Monkman Shelley, Kirpalani Haresh
The Hospital for Sick Children, Toronto, Canada.
Resuscitation. 2009 Jan;80(1):120-5. doi: 10.1016/j.resuscitation.2008.08.021. Epub 2008 Oct 25.
High oxygen increases morbidity and mortality. Current guidelines in Neonatal Resuscitation Programme (NRP) state if self-inflating bags are used with an input FiO2 of 1.0 without an oxygen reservoir a delivered safe FiO2 of approximately 0.40 is achieved. This conflicts with manufacturer findings (Laerdal infant resuscitator (LIR)). We assessed FiO2 delivery by the LIR, varying oxygen reservoir (OR) use, ventilation and input flowrates.
A test lung was connected to the LIR and oxygen analyzer. FiO2 delivery was measured under these four conditions: LIR plus OR and FiO2 1.0 or FiO2 0.40; LIR minus OR and FiO2 of 1.0 and 0.40. Variations of ventilation patterns in random order, assessed tidal volumes (from 20 and 40mL), ventilation rates (from 30, 40 and 60breaths/min), and input flowrates (from 1, 3, 5, 8, and 10Lpm). A wash-out period of 1min of ventilation was followed by measure of FiO2 during manual ventilation.
The measured FiO2 with the LIR delivered the same source FiO2 under all experimental conditions for flowrates of 5Lpm and greater; irrespective of the OR presence or absence. At flowrates of 1 and 3Lpm, FiO2 was lower, with and without the reservoir, but the reservoir was visibly identified as not filled.
Our findings support the manufacturers performance specification that high input FiO2 results in high delivered FiO2 with/without an OR. These results dispute the 2006 NRP guidelines that state "in the absence of a reservoir (oxygen) the delivered oxygen is reduced to about 40%".
高氧会增加发病率和死亡率。新生儿复苏计划(NRP)的现行指南指出,如果使用自充气式气囊且输入的氧浓度(FiO2)为1.0且没有氧气储存器,则可实现约0.40的安全输送FiO2。这与制造商的研究结果(Laerdal婴儿复苏器(LIR))相冲突。我们评估了LIR在不同氧气储存器(OR)使用情况、通气和输入流速下的FiO2输送情况。
将一个测试肺连接到LIR和氧分析仪上。在以下四种条件下测量FiO2输送情况:LIR加OR且FiO2为1.0或FiO2为0.40;LIR减去OR且FiO2为1.0和0.40。以随机顺序改变通气模式,评估潮气量(20和40毫升)、通气频率(30、40和60次/分钟)以及输入流速(1、3、5、8和10升/分钟)。在手动通气期间,先进行1分钟的通气洗脱期,然后测量FiO2。
对于5升/分钟及更高的流速,在所有实验条件下,LIR测得的FiO2与源FiO2相同;无论是否存在OR。在1和3升/分钟的流速下,无论有无储存器,FiO2都较低,但明显可看出储存器未充满。
我们的研究结果支持制造商的性能规格,即无论有无OR,高输入FiO2都会导致高输送FiO2。这些结果与2006年NRP指南中“在没有(氧气)储存器的情况下,输送的氧气会降至约40%”的说法相矛盾。