Newton C M
Bell Laboratories, 600 Mountain Avenue, Murray Hill, New Jersey 07974, U.S.A.
Talanta. 1977 Jun;24(6):377-81. doi: 10.1016/0039-9140(77)80023-4.
High-precision gravimetric coulometry with a silver-perchloric acid coulometer is evaluated as an alternative to the conventional titrimetric method. The loss of weight (caused by electrolytic dissolution) of a highly pure silver anode in series with the cathode of a conventional constant-current titration system is measured and related to the number of equivalents of substance titrated. The precision of the method is determined by titrations of the Standard Reference Material 83C arsenious oxide (99.99% pure) with electrogenerated iodine, using biamperometric end-point detection. Depending on the size of the sample, an ultimate precision of 25 ppm is obtained. The assay for 0.5-g samples of the SRM material is 99.993(9) +/- 0.002(5)% purity.
对使用高氯酸银电量计的高精度重量库仑法进行了评估,以替代传统的滴定法。测量与传统恒流滴定系统阴极串联的高纯度银阳极的失重(由电解溶解引起),并将其与被滴定物质的当量数相关联。该方法的精密度通过使用双安培终点检测法,用电生碘滴定标准参考物质83C三氧化二砷(纯度99.99%)来确定。根据样品大小,可获得25 ppm的最终精密度。对0.5 g SRM材料样品的分析纯度为99.993(9) +/- 0.002(5)%。