Abstract

A new polymer matrix based membrane electrode with an ion-exchanger responding to calcium was constructed by dissolving the copolymer ethylene-vinyl-acetate together with the ion-exchanger in chloroform in the presence of a mixture of dioctylphthalate-nitrobenzene as plasticizer. The ion-exchanger used as the electroactive component was calcium didecyl phosphate in di-(n-octylphenyl) phosphonate (Orion). This electrode exhibited near-Nernstian response over the concentration range 10(-1)-4 x 10(-6)M calcium. The pH did not affect the electrode performance within the range 8-11. Response time varied from 15 to 120 sec and the lifetime exceeded six months. The membrane is subject to static charge buildup, but this is avoided by controlling the level of dryness of the membrane. Selectivity coefficients determined for both monovalent and divalent cations showed negligible interference by most of these ions. The electrode was applied successfully to the determination of calcium in commercial mineral waters.

摘要

通过在邻苯二甲酸二辛酯 - 硝基苯混合物作为增塑剂存在的情况下,将共聚物乙烯 - 醋酸乙烯酯与离子交换剂溶解于氯仿中,构建了一种基于新型聚合物基质的、带有对钙有响应的离子交换剂的膜电极。用作电活性成分的离子交换剂是二(正辛基苯基)膦酸酯(奥立龙)中的磷酸二癸酯钙。该电极在10⁻¹ - 4×10⁻⁶M钙的浓度范围内呈现近能斯特响应。在8 - 11的pH范围内,pH不影响电极性能。响应时间在15到120秒之间变化,使用寿命超过六个月。该膜会产生静电荷积累,但通过控制膜的干燥程度可避免这种情况。对单价和二价阳离子测定的选择性系数表明,大多数这些离子的干扰可忽略不计。该电极已成功应用于商业矿泉水中钙的测定。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验