Abstract

Properties of six derivatives of 1-trifluoroacetylbenzene: [4-(n-butyl)- (1), 4-(n-hexadecyl)- (2), 4-dodecyloxy- (3), 4-(n-dodecylsulfonyl)- (4), N,N-dioctyl-4-trifluoroacetylbenzamide (5), octyl-p-trifluoroacetylbenzoate (6)] as neutral carriers for carbonate ion were examined and compared. The sensitivity towards carbonate ion was for (3) pH dependent. This eliminates (3) from practical applications in clinical analysis. When measuring CO(3)(2-) within the physiological range of human blood using as carriers compounds 1 and 2 the interference of chloride must be taken into account. In the case of carriers 4, 5, 6 this effect is negligible. Electrodes with membranes containing as carriers 2, 4, 5 and 6 were tested in an automatic potentiometric clinical analyser Microlyte 6, KONE. To avoid contamination by atmospheric CO(2) of three aqueous standards (TES, NaCl, NaHCO(3)), pH was adjusted by coulomeric generation of H(+) or OH(-) in a system devoid of carbon dioxide. Recovery of HCO(3)(-) calculated from measured CO(3)(2-) and pH, was investigated in a series of aqueous solutions and spiked bovine serum samples. The correlation between added and recovered concentration of HCO(3)(-) was linear with the intercept close to 0 and slope equal to 1 in aqueous solutions for all ligands and in bovine serum samples only in the case of ligand (2).

摘要

1-三氟乙酰苯的六种衍生物[4-(正丁基)-(1)、4-(正十六烷基)-(2)、4-十二烷氧基-(3)、4-(正十二烷基磺酰基)-(4)、N,N-二辛基-4-三氟乙酰苯甲酰胺(5)、辛基对三氟乙酰苯甲酸酯(6)]作为碳酸根离子中性载体的性能进行了研究和比较。对碳酸根离子的敏感性对于(3)来说取决于pH值。这使得(3)无法用于临床分析的实际应用中。当使用化合物1和2作为载体在人体血液的生理范围内测量CO₃²⁻时,必须考虑氯离子的干扰。对于载体4、5、6,这种影响可以忽略不计。含有载体2、4、5和6的膜电极在自动电位临床分析仪Microlyte 6(科纳公司)中进行了测试。为避免三种水溶液标准液(TES、NaCl、NaHCO₃)被大气中的CO₂污染,在一个不含二氧化碳的系统中通过库仑法产生H⁺或OH⁻来调节pH值。在一系列水溶液和加标的牛血清样品中研究了根据测得的CO₃²⁻和pH值计算出的HCO₃⁻的回收率。对于所有配体,在水溶液中以及仅在配体(2)的情况下在牛血清样品中,添加的HCO₃⁻浓度与回收浓度之间的相关性呈线性,截距接近0且斜率等于1。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验