Acedo-Valenzuela María-Isabel, Galeano-Díaz Teresa, Mora-Díez Nielene, Silva-Rodríguez Antonio
Departamento de Química Analítica, Facultad de Ciencias, Universidad de Extremadura, Av. de Elvas s/n, 06071-Badajoz, Spain.
Talanta. 2005 May 15;66(4):952-60. doi: 10.1016/j.talanta.2004.12.044. Epub 2005 Jan 21.
A semiautomatic extraction-fluorimetric method for the determination of tricyclic antidepressant drugs (TCAs) based in the formation of ion pairs with 9,10-dimethoxyanthracene-2-sulphonate (DMAS) has been developed. The aqueous solutions of the TCAs (imipramine, desipramine, amitriptyline, nortriptyline, clomipramine or doxepine) are injected into a carrier composed by DMAS in an acid medium and the ion pair formed is extracted into dichloromethane where the fluorescence is measured. An experimental design (Central Composite Design) together with the Response Surface Methodology has been used to find the optimal instrumental FIA and chemical variables. We have considered as the response function the product of the peak height by the sampling frequency. The calibration curves were linear over the working range (0.25-3.00mgL(-1)). The limits of detection were lower than 0.30mgL(-1). The method has been satisfactorily applied to the determination of imipramine, amitriptyline, clomipramine and doxepin in pharmaceutical preparations.
已开发出一种半自动萃取荧光法,用于测定三环类抗抑郁药(TCA),该方法基于与9,10 - 二甲氧基蒽 - 2 - 磺酸盐(DMAS)形成离子对。将TCA(丙咪嗪、地昔帕明、阿米替林、去甲替林、氯米帕明或多塞平)的水溶液注入由DMAS在酸性介质中组成的载体中,形成的离子对被萃取到二氯甲烷中,然后在其中测量荧光。采用实验设计(中心复合设计)和响应面方法来寻找最佳的仪器流动注射分析(FIA)和化学变量。我们将峰高与采样频率的乘积作为响应函数。校准曲线在工作范围(0.25 - 3.00mgL⁻¹)内呈线性。检测限低于0.