Wen Yi, Wang Ying, Feng Yu-Qi
Department of Chemistry, Wuhan University, Wuhan 430072, PR China.
Talanta. 2006 Aug 15;70(1):153-9. doi: 10.1016/j.talanta.2005.11.049. Epub 2006 Jan 6.
An on-line simple and rapid method for the simultaneous determination of tetracycline (TC), oxytetracycline (OTC), chlortetracycline (CTC) and doxycycline (DC) residues in fish muscle was developed by coupling in-tube solid-phase microextraction (SPME) to high-performance liquid chromatography (HPLC) with a photodiode array detector. Biocompatible poly (methacrylic acid-ethylene glycol dimethacrylate) monolithic capillary was selected as the extraction medium, and no precipitating protein and removing fat steps were required prior to extraction. In order to optimize the extraction of these compounds, several in-tube SPME parameters were investigated. Simply performed by extracting with 0.01M EDTA-MacIlvaine buffer solution (pH 4.0) and centrifugation, the sample then could be directly injected into the device for extraction. The limits of detection of tetracycline, oxytetracycline, chlortetracycline and doxycycline were calculated to be 22, 16, 30 and 21ng/g, respectively. The calibration curves showed linearity in the range of 100-10,000ng/g with a linear coefficient R(2) value above 0.9980. Excellent method reproducibility was found by intra- and inter-day precisions, yielding the R.S.D.s less than 4.22% and 5.71%, respectively.