Salomons G J, Kerr A T, Mei X, Patel D
Department of Medical Physics, Cancer Centre of South Eastern Ontario, 25 King Street, Kingston, Ontario K7L 5P9, Canada.
Med Phys. 2008 Oct;35(10):4289-91. doi: 10.1118/1.2975145.
In November 2007 Varian Medical Systems released a new version of ECLIPSE (version 8.1.17), which includes a correction to its Anisotropic Analytical Algorithm affecting monitor unit calculations for Enhanced Dynamic Wedge. The purpose of this study is to evaluate and compare the accuracy of the MU calculations for version 8.1.17 with that from the previous version (8.0.05).
2007年11月,瓦里安医疗系统公司发布了新版的ECLIPSE(8.1.17版),其中对其各向异性分析算法进行了修正,该修正影响了增强动态楔形板的监测单位计算。本研究的目的是评估并比较8.1.17版与上一版(8.0.05版)监测单位计算的准确性。