Reiher Alexandra E, Gallagher Thomas A, Emanuele Mary Ann
Division of Endocrinology, Loyola University Medical Center, Maywood, Illinois 60153, USA.
Endocrine. 2009 Feb;35(1):22-4. doi: 10.1007/s12020-008-9122-6. Epub 2008 Dec 13.
Insulin pumps operate almost universally in the work environment. Tubeless insulin pumps are a relatively recent development for management of insulin-dependent diabetes. This is the first report of location-dependent failure of insulin pump function, possibly due to external interference, occurring only in the setting of a medical center. A successful solution to this problem was designed and implemented. In conclusion, heightened awareness of this possibility in individuals who use tubeless insulin pumps in a medical environment would seem prudent, as when identified, it can be easily corrected.