使用多电极消融导管通过占空比双极和单极射频能量进行肺静脉隔离。

Pulmonary vein isolation by duty-cycled bipolar and unipolar radiofrequency energy with a multielectrode ablation catheter.

作者信息

Boersma Lucas V A, Wijffels Maurits C E F, Oral Hakan, Wever Eric F D, Morady Fred

机构信息

Cardiology Department, St. Antonius Hospital, Nieuwegein, The Netherlands.

出版信息

Heart Rhythm. 2008 Dec;5(12):1635-42. doi: 10.1016/j.hrthm.2008.08.037. Epub 2008 Sep 4.

Abstract

BACKGROUND

Pulmonary vein (PV) isolation for ablation of atrial fibrillation (AF) remains a complex and lengthy procedure.

OBJECTIVE

The purpose of this study was to evaluate the feasibility and safety of a novel multielectrode catheter that delivers duty-cycled bipolar and unipolar radiofrequency (RF) energy.

METHODS

Patients eligible for catheter ablation of paroxysmal AF after screening with magnetic resonance imaging and transesophageal echocardiography were included in the study. A decapolar (3-mm electrode, 3-mm spacing, 25-mm diameter), circular, over-the-wire mapping and ablation catheter was deployed in the antrum of each PV. Ablation was performed with 60-second, 60 degrees C applications of duty-cycled bipolar/unipolar RF in a 4:1 ratio simultaneously at all selected electrode pairs until local activity was no longer observed. At 6 months, 7-day Holter monitoring was performed to determine freedom from AF without use of antiarrhythmic drugs.

RESULTS

In 98 patients (mean age 59 +/- 9 years), the PV ablation catheter was used for ablation of 369 veins (20 common left antra). All targeted veins (100%) were isolated as confirmed by the absence of potentials in the ostium either by PV ablation catheter or Lasso mapping. Mean number of RF applications was 27 +/- 7, total procedural time 84 +/- 29 minutes, and fluoroscopy time 18 +/- 8 minutes. Follow-up after 6 months without antiarrhythmic drugs showed freedom from AF in 83% of patients. No procedure-related complications were observed.

CONCLUSION

PV isolation by duty-cycled bipolar/unipolar low-power RF energy through a circular, decapolar catheter can be achieved safely and efficiently, with good efficacy at 6 months.

摘要

背景

用于心房颤动(AF)消融的肺静脉(PV)隔离仍是一个复杂且耗时的过程。

目的

本研究旨在评估一种新型多电极导管的可行性和安全性,该导管可输送占空比控制的双极和单极射频(RF)能量。

方法

经磁共振成像和经食管超声心动图筛查符合阵发性AF导管消融条件的患者纳入本研究。将一个十极(电极3毫米、间距3毫米、直径25毫米)圆形、可过导丝的标测和消融导管置于每根PV的窦部。在所有选定的电极对处以4:1的比例同时进行占空比控制的双极/单极RF 60秒、60℃的应用,直至不再观察到局部活动。在6个月时,进行7天动态心电图监测以确定在不使用抗心律失常药物的情况下无AF。

结果

在98例患者(平均年龄59±9岁)中,PV消融导管用于消融369条静脉(20个共同的左窦)。通过PV消融导管或Lasso标测确认,所有目标静脉(100%)均被隔离,窦口无电位。平均RF应用次数为27±7次,总手术时间84±29分钟,透视时间18±8分钟。6个月不使用抗心律失常药物的随访显示,83%的患者无AF。未观察到与手术相关的并发症。

结论

通过圆形十极导管利用占空比控制的双极/单极低功率RF能量进行PV隔离可安全、有效地实现,6个月时疗效良好。